  • Question 1 - What feature is found in the flowmeters of modern anaesthetic machines ? ...

    Incorrect

    • What feature is found in the flowmeters of modern anaesthetic machines ?

      Your Answer: The reading of the flowmeter is taken from the midpoint of the bobbin

      Correct Answer: The use of stannic oxide increases the accuracy of flowmeters

      Explanation:

      Flowmeters measure the rate at which a specific gas, that the flowmeter has been calibrated for, passes through. This calibration is done at room temperature and standard atmospheric pressure with an accuracy of +/- 2%.

      Reading the flowmeter is done from the top of a bobbin (the midpoint of a ball). Oxygen is the last gas to be added downstream to the mixture delivered to the back bar as a safety feature. This prevents delivery of a hypoxic mixture.

      Inaccurate flow measurements occur when the bobbin sticks to the inside wall of the flowmeter. Stannic oxide has been used as a successful antistatic substance thus, reducing the aforementioned risk.

      Carbon dioxide being easily delivered is found on some older machines, but those attached flowmeters are limited by a maximum flow of 500 ml /min. Thus avoiding the delivery of a hypercarbic mixture.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      29.6
      Seconds
  • Question 2 - Which of the following is correct for gas pipeline pressure? ...

    Incorrect

    • Which of the following is correct for gas pipeline pressure?

      Your Answer: 7 bar for Entonox

      Correct Answer: 4 bar for oxygen

      Explanation:

      Pipeline gases (in the UK this includes: Oxygen, Nitrous oxide, Medical air, and Entonox) are supplied at 4 bar (or 400 kPa), and compressed air is supplied at 7 bar for power tools.

      Carbon dioxide and nitric oxide are usually only supplied in cylinders.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      35.8
      Seconds
  • Question 3 - Regarding laminar gas flow, which of the following options has the most influence...

    Correct

    • Regarding laminar gas flow, which of the following options has the most influence on laminar flow?

      Your Answer: Diameter of tube

      Explanation:

      Laminar flow can be defined as the motion of a fluid where every particle in the fluid follows the same path of its previous particles. The following are properties of laminar flow of gas or fluids:

      1. Smooth unobstructed flow of gas through a tube of relatively uniform diameter
      2. Few directional changes
      3. Slow, steady flow through straight smooth, rigid, large calibre, cylindrical tube
      4. Outer layer flow slower than the centre due to friction, results in discrete cylindrical layers, or streamlines
      5. Double flow by doubling pressure as long as the flow pattern remains laminar

      Poiseuille’s Law relates the factors that determine laminar flow. It indicates the degree of resistance to fluid flow through a tube. The resistance to fluid flow through a tube is directly related to the length, flow and viscosity; and inversely related to the radius of the tube to the fourth power. This means that, when the radius is doubled, there is increase in flow by a factor of 16.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      9.2
      Seconds
  • Question 4 - A 74-year-old with a VVI pacemaker is undergoing a hip replacement. Which of...

    Correct

    • A 74-year-old with a VVI pacemaker is undergoing a hip replacement. Which of the following is most likely to predispose him to an electrical hazard?

      Your Answer: Use of cutting unipolar diathermy

      Explanation:

      A single chamber pacemaker was implanted in the patient. In VVI mode, a pacemaker paces and senses the ventricle while being inhibited by a perceived ventricular event. The most likely electrical hazard from diathermy is electromagnetic interference (EMI).

      EMI has the potential to cause the following: Inhibition of pacing
      Asynchronous pacing
      Reset to backup mode
      Myocardial burns, and
      Trigger VF.

      Diathermy entails the implementation of high-frequency electrical currents to produce heat and either make incisions or induce coagulation. Monopolar cautery involves disposable cautery pencils and electrosurgical diathermy units. In typical monopolar cautery, an electrical plate is placed on the patient’s skin and acts as an electrode, while the current passes between the instrument and the plate. Monopolar diathermy can therefore interfere with implanted metal devices and pacemaker function.

      Bipolar diathermy, where the current passes between the forceps tips and not through the patient and is less likely to generate EMI.

      Whilst the presence of a CVP line may in theory predispose the patient to microshock, the use of prerequisite CF electrical equipment makes this very unlikely. The presence of a CVP line and pacemaker does not therefore unduly increase the risk of an electrical hazard.

      Isolating transformers are used to protect secondary circuits and individuals from electrical shocks. There is no step-up or step-down voltage (i.e. there is a ratio of 1 to 1 between the primary and secondary windings).

      A ground (or earth) wire is normally connected to the metal case of an operating table to protect patients from accidental electrocution. In the event that a fault allows a live wire to make contact with the metal table (broken cable, loose connection etc.) it becomes live. The earth will provide an immediate path for current to safely flow through and so the table remains safe to touch. Being a low resistance path, the earth lets a large current flow through it when the fault occurs ensuring that the fuse or RCD will quickly blow. Without an operating table earth, the patient is not at more risk of an electrical hazard because of the pacemaker.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      23.8
      Seconds
  • Question 5 - A 240 volt alternating current (AC) socket from a wall is used to...

    Incorrect

    • A 240 volt alternating current (AC) socket from a wall is used to charge a direct current (DC) cardiac defibrillator. Name the electrical component that converts AC to DC.

      Your Answer: Inductor

      Correct Answer: Rectifier

      Explanation:

      There are two types of defibrillators
      AC defibrillator
      DC defibrillator

      AC defibrillator,
      consists of a step-up transformer with primary and secondary winding and two switches. Since secondary coil consists of more turns of wire than the primary coil, it induces larger voltage. A voltage value ranging between 250V to 750V is applied for AC external defibrillator. And used to enable the charging of a capacitor.

      DC defibrillator,
      consists of auto transformer T1 that acts as primary of the high voltage transformer T2. Is an iron core that transfers energy between 2 circuits by electromagnetic induction. Transformers are used to isolate circuits, change impedance and alter voltage output. transformers do not convert AC to DC.

      Diode rectifier composed of 4 diodes made of semiconductor material allows current to flow only in one direction. Alternating current (AC) passing through these diodes produces direct current (DC). Capacitor stores the charge in the form of an electrostatic field.

      Capacitor is used to convert the rectified AC voltage to produce DC voltage but capacitors do not directly convert AC to DC.

      Inductor induces a counter electromotive force(emf) that reduces the capacitor discharge value.

      In step-down transformer primary coils has more turns of wire than secondary coil, so induced voltage is smaller in the secondary coil.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      10.2
      Seconds
  • Question 6 - A strain gauge operates on which of the following principles? ...

    Incorrect

    • A strain gauge operates on which of the following principles?

      Your Answer: Uncoiling of a tube by pressurised gas

      Correct Answer: Varying resistance of a wire when stretched

      Explanation:

      The principle by which a strain gauge works is that when a wire is stretched, it becomes longer and thinner, and as a result, its resistance increases.

      A strain gauge, which is used in pressure transducers, acts as a resistor. When the pressure in a pressure transducer changes, the diaphragm moves, changing the tension in the resistance wire and thus changing the resistance.

      Changes in current flow through the resistor are amplified and displayed as a pressure change measure.

      A Wheatstone bridge, on the other hand, is frequently used to measure or monitor these changes in resistance.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      8.5
      Seconds
  • Question 7 - Regarding anaesthetic breathing circuits, which one would be the best for spontaneous breathing?...

    Incorrect

    • Regarding anaesthetic breathing circuits, which one would be the best for spontaneous breathing?

      Your Answer: Mapleson E

      Correct Answer: Lack circuit

      Explanation:

      Among the breathing circuits, the Lack circuit is the most efficient for spontaneous breathing.

      An outer coaxial tube is present to deliver fresh air; exhaust air is routed to an inner tube, which is then delivered to a scavenging system. An expiratory valve is seen at the patient end, which is an advantage over other circuits. Moreover, the Lack circuit prevents rebreathing slightly greater than the alveolar minute ventilation at 4-5 litres per minute.

      The Bain circuit prevents rebreathing at 160-200ml/kg per minute, and is a co-axial version of the Mapleson D circuit.

      The Mapleson E circuit prevent rebreathing at a fresh gas flow (FGF) of approximately twice the patient’s normal minute volume. A modification of this, the Mapleson F, has a reservoir bag at the opposite end for the FGF. This circuit is appropriate for paediatric patients with a body weight less than 20 kg.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      28.5
      Seconds
  • Question 8 - Prior to rapid sequence induction of anaesthesia, a man with a BMI of...

    Correct

    • Prior to rapid sequence induction of anaesthesia, a man with a BMI of 35 is pre-oxygenated. Which method of pre-oxygenation with a tight-fitting face mask is the most effective?

      Your Answer: Oxygen 6 litres per minute via a Mapleson A breathing system, with patient sitting up at 30 degrees breathing four vital capacity breaths

      Explanation:

      This patient is morbidly obese and has a high risk of developing hypoxia. This will be exacerbated by the patient’s supine position, as a result of:

      Functional residual capacity has been reduced (FRC)
      Increased closing capacity (CC)
      Reduced tidal volume due to increased airway resistance, decreased thoracic cage compliance, and decreased respiratory muscle strength and endurance
      Following induction of general anaesthesia, there is a tendency for atelectasis and increased O2 consumption due to the increased workload of respiratory muscles and the overall increase in metabolism.

      Pre-oxygenation with 100 percent oxygen via a tight-fitting mask can be done using either tidal volume breaths for three to five minutes or four vital capacity breaths in normal circumstances. In the head-up position, this patient is much more likely to be adequately pre-oxygenated, maximising the FRC and minimising the CC. In spontaneously breathing patients, the Mapleson A and circle systems are both effective, but the Mapleson D requires 160-200 ml/kg/minute to prevent rebreathing.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      56.3
      Seconds
  • Question 9 - Which of the following statements is true about oxygen face masks? ...

    Incorrect

    • Which of the following statements is true about oxygen face masks?

      Your Answer: Rebreathing of alveolar gas is a feature associated with fixed performance masks

      Correct Answer: The oxygen concentration delivered by high air flow oxygen enrichment devices is not dependent on the respiratory pattern of the patient

      Explanation:

      The normal peak inspiratory flow in healthy individuals is 20-30 L/min during each normal tidal ventilation. This is expected to increase with greater respiratory rate and deeper inspiration.

      Face masks are used to facilitate the delivery of oxygen from a breathing system to a patient. Face masks can be divided into two types: fixed performance or variable performance devices.

      In fixed performance devices (also known as high air flow oxygen enrichment or HAFOE), fixed inspired oxygen concentration is delivered to the patent, independent and greater than that of the patient’s peak inspiratory flow rate (PIFR). No random entrainment is expected to occur at the time of PIFR, hence, the oxygen concentration in HAFOE devices is not dependent on the patient’s respiratory pattern.

      Moreover, in HAFOE masks, the concentration of oxygen at a given oxygen flow rate is determined by the size of the constriction; a device with a greater entrainment aperture delivers a lower oxygen concentration. Therefore, a 40% Venturi device will have lesser entrainment aperture when compared to a 31% Venturi. Venturi masks allow relatively fixed concentrations of supplemental oxygen to be inspired e.g. 24%, 28%, 31%, 35%, 40% and 60% oxygen. These are colour coded and marked with the recommended oxygen flow rate.

      Variable performance devices deliver variable inspired oxygen concentration to the patient, and is dependent on the PIFR. The PIFR can often exceed the flow rate at which oxygen or an oxygen/air mixture is supplied by the device, depending on a patient’s inspiratory effort. In addition, these masks allow expired air to be released through the holes in the sides of the mask. Thus, with increased respiratory rate, rebreathing of alveolar gas from inside the mask may occur.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      4889.4
      Seconds
  • Question 10 - The Control of Substances Hazardous to Health (COSHH) regulations recommend air supply rates...

    Incorrect

    • The Control of Substances Hazardous to Health (COSHH) regulations recommend air supply rates to specific environments. Which of the following statements is true?

      Your Answer: Anaesthetic rooms receive a volume of 0.65 m3 of air per second

      Correct Answer: Preparation rooms receive a volume of 0.1 m3 of air per second

      Explanation:

      Control of Substances Hazardous to Health (COSHH) was established by government under the Health and Safety at Work act in 1989. Their employers work on identification and management of those substances that are dangerous to health. The implications for anaesthetists include gas scavenging, equipment contamination and environmental safety. Adequate ventilation is required in areas where anaesthetic gases are present. The minimum air supply that is legally required in each specific area is: Operating theatres: 0.65 m3/second. Anaesthetic rooms: 0.15 m3/s. Preparation rooms: 0.1 m3/s. Recovery rooms need 15 air changes per hour.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      32.3
      Seconds
  • Question 11 - Bioelectric potentials that have been measured have an optimum bandwidth and typical frequency....

    Incorrect

    • Bioelectric potentials that have been measured have an optimum bandwidth and typical frequency. For a standard 12-lead ECG, which of the following bandwidth and voltage combinations is the best?

      Your Answer: Bandwidth 1-5000 Hz, voltage 10-200 microvolts

      Correct Answer: Bandwidth 0.05-150 Hz, voltage 100-4000 microvolts

      Explanation:

      The potential difference (amplitude) and bandwidth frequencies of bioelectric signals are typical.

      These are the following:

      ECG: A bandwidth of 0.5-50 Hz is usually sufficient in monitoring mode, but a typical diagnostic bandwidth is 0.05-150 Hz (up to 200 Hz) with a typical voltage range of 0.1-4 millivolts (100-4000 microvolts).
      EEG has a frequency range of 0.5-100 Hz and a voltage range of 0.5-100 microvolts.
      EMG has a frequency range of 0.5 to 350 Hz and a voltage range of 0.5 to 30 millivolts.

      Prior to display, these small signals will need to be amplified and processed further.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      46
      Seconds
  • Question 12 - An emergency appendicectomy is being performed on a 20 year old man. For...

    Incorrect

    • An emergency appendicectomy is being performed on a 20 year old man. For maintenance of anaesthesia, he is being ventilated using a circle system with a fresh gas flow (FGF) of 1 L/min (air/oxygen and sevoflurane). The trace on the capnograph shows a normal shape. The table below demonstrates the changes in the end-tidal and baseline carbon dioxide measurements of the capnograph at 10 and 20 minutes of anaesthesia maintenance. End-tidal CO2: 4.9 kPa vs 8.4kPa (10 minutes vs 20 minutes). Baseline end-tidal CO2: 0.2 kPa vs 2.4kPa. Pulse 100-107 beats per minute, systolic blood pressure 125-133 mmHg and oxygen saturation 98-99%. Which of the following is the single most important immediate course of action?

      Your Answer: Check the patient's core temperature

      Correct Answer: Increase the FGF

      Explanation:

      End-tidal carbon dioxide (ETCO2) monitoring has been an important factor in reducing anaesthesia-related mortality and morbidity. Hypercarbia, or hypercapnia, occurs when levels of CO2 in the blood become abnormally high (Paco2 >45 mm Hg). Hypercarbia is confirmed by arterial blood gas analysis. When using capnography to approximate Paco2, remember that the normal arterial€“end-tidal carbon dioxide gradient is roughly 5 mm Hg. Hypercarbia, therefore, occurs when PETco2 is greater than 40 mm Hg.

      The most likely explanation for the changes in capnograph is either exhaustion of the soda lime and a progressive rise in circuit dead space.

      Inspect the soda lime canister for a change in colour of the granules. To overcome soda lime exhaustion, the first step is to increase the fresh gas flow (FGF) (Option A). Then, if need arises, replace the soda lime granules. Other strategies that can work are changing to another circuit or bypassing the soda lime canister, but remember that both these strategies are employed only after increasing FGF first. Exclude other causes of equipment deadspace too.

      There are also other causes for hypercarbia to develop intraoperatively:
      1. Hypoventilation is the most common cause of hypercapnia. A. Inadequate ventilation can occur with spontaneous breathing due to drugs like anaesthetic agents, opioids, residual NMDs, chronic respiratory or neuromuscular disease, cerebrovascular accident.
      B. In controlled ventilation, hypercapnia due to circuit leaks, disconnection or miscalculation of patient’s minute volume.
      2. Rebreathing – Soda lime exhaustion with circle, inadequate fresh gas flow into Mapleson circuits and increased breathing system deadspace.
      3. Endogenous source – Tourniquet release, hypermetabolic states (MH or thyroid storm) and release of vascular clamps.
      4. Exogenous source – Absorption of CO2 from pneumoperitoneum.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      153.8
      Seconds
  • Question 13 - Many of the processes we deal with in anaesthesia entail a relationship between...

    Incorrect

    • Many of the processes we deal with in anaesthesia entail a relationship between two or more variables. Which of the following relationships is a linear relationship?

      Your Answer: The relationship between the pressure of a fixed mass of gas and its volume at a fixed temperature

      Correct Answer: The relationship between the junction potential and temperature in a thermocouple

      Explanation:

      Two bonded wires of dissimilar metals, iron/constantan or copper/constantan, make up a thermocouple (constantan is an alloy of copper and nickel). At the tip, a thermojunction voltage is generated that is proportional to temperature (Seebeck effect).

      All of the other connections are non-linear.

      For a single compartment model, the relationship between a decrease in plasma concentration of an intravenous bolus of a drug and time is a washout exponential.

      A sine wave is the relationship between current and degrees or time from a mains power source.

      A sigmoid curve represents the relationship between efficacy and log-dose of a pure agonist on mu receptors.

      The pressure of a fixed mass of gas and its volume (Boyle’s law) at a fixed temperature are inversely proportional, resulting in a hyperbolic curve.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      47.4
      Seconds
  • Question 14 - International colour coding is used on medical gas cylinders. Other characteristics also play...

    Correct

    • International colour coding is used on medical gas cylinders. Other characteristics also play a role in determining the gas's identity within a cylinder. Which of the following options best describes a cylinder containing analgesics for obstetrics?

      Your Answer: Blue body, blue/white shoulder, full cylinder; 13700 KPa, gas mixture, requires a dual stage pressure regulator

      Explanation:

      The body of the Entonox cylinder is usually blue (occasionally white), with blue and white shoulders. Entonox contains a 50:50 mixture of oxygen and nitrous oxide, with a full cylinder pressure of 13700 KPa (137 bar). The cylinder is equipped with a two-stage pressure regulator for safe operation.

      The cylinder body and shoulder of nitrous oxide are (French) blue.

      In today’s anaesthetic workstations, carbon dioxide cylinders are no longer used.

      The body of an oxygen cylinder is black, with a white shoulder.

      The white Heliox (21 percent oxygen and 79 percent helium) cylinder has a brown and white shoulder. The administration of this gas mixture, which is less dense than air, is used to reduce turbulence (stridor) of inspiratory flow in patients with upper airway obstruction.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      23.6
      Seconds
  • Question 15 - Regarding the emergency oxygen flush, which is true? ...

    Correct

    • Regarding the emergency oxygen flush, which is true?

      Your Answer: May lead to awareness if used inappropriately

      Explanation:

      When the emergency oxygen flush is pressed, 100% oxygen is supplied from the common gas outlet. This gas bypasses BOTH flowmeters and vaporisers. The flow of oxygen is usually 45 l/min at a PRESSURE OF 400 kPa.

      There is an increased risk of pulmonary barotrauma when the emergency flush is pressed, especially when anaesthetising paediatric patients.

      The inappropriate use of the flush causes dilution of anaesthetic gases and this increases the possibility of anaesthetic awareness.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      31.7
      Seconds
  • Question 16 - All of the following statements about pH electrode are incorrect except: ...

    Correct

    • All of the following statements about pH electrode are incorrect except:

      Your Answer: A semi-permeable membrane reduces protein contamination

      Explanation:

      Pulse oximeters combine the principles of oximetry and plethysmography to noninvasively measure oxygen saturation in arterial blood. A sensor containing two or three light emitting diodes and a photodiode is placed across a perfused body part, commonly a finger, to be transilluminated. Oximetry depends on oxyhaemoglobin and deoxyhaemoglobin, and their ability to absorb the beams of light produced by the light emitting diodes: red light at 660 nm and infrared light at 960 nm.

      The isosbestic point is the point wherein two different substances absorb light to the same extent. For oxyhaemoglobin and deoxyhaemoglobin, the points are at 590 nm and 805 nm. These are considered reference points where light absorption is independent of the degree of saturation.

      Non-constant absorption of light is often due to the presence of an arterial pulsation, whilst constant absorption of light is seen in non-pulsatile tissues.

      Most pulse oximeters are inaccurate at low SpO2, but is accurate at +/- 2% within the range of 70% to 100% SpO2. All pulse oximeters demonstrate a delay in between changes in SaO2 and SpO2, and display average readings every 10 to 20 seconds, hence they are unable to detect acute desaturation episodes.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      26.5
      Seconds
  • Question 17 - Do the medical gas cylinders in the UK have the following colour codes?...

    Incorrect

    • Do the medical gas cylinders in the UK have the following colour codes?

      Your Answer: Carbon dioxide cylinders are brown

      Correct Answer: Oxygen cylinders have a black body with white shoulders

      Explanation:

      The following are the colour codes for medical gas cylinder in the UK:

      Oxygen cylinder has a dark body with white shoulders.

      Nitrous oxide is French blue. Air encompasses a grey body with dark and white quarters on the shoulders.

      Entonox contains a French blue body with white and blue quarters on the shoulders.

      Carbon dioxide barrels are grey in colour.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      143.3
      Seconds
  • Question 18 - A measuring system's response to change is complex, yet it can be mathematically modelled. Which of the following terms best characterises a pressure transducer's responsiveness to blood pressure changes? ...

    Incorrect

    • A measuring system's response to change is complex, yet it can be mathematically modelled. Which of the following terms best characterises a pressure transducer's responsiveness to blood pressure changes?

      Your Answer:

      Correct Answer: Dynamic second-order response

      Explanation:

      The static-response defines how a measuring system behaves while it is in equilibrium (i.e. when the measured values are not changing). If the value being measured changes over time, the reaction of a measuring system will change as well which would be a dynamic response.
      The dynamic response of a measuring system can be subdivided into zero-order, first-order and second-order responses:

      Zero-order:
      Consider a thermometer that has been left in a room for a week. The thermometer will display the current ambient temperature when you enter the room.

      First-order:
      Consider the use of a mercury thermometer to check a patient’s temperature. It is comprised of a mercury column that expands as it warms up. The scale’s initial temperature is room temperature, but when it’s placed under the patient’s tongue, the temperature readings rise until they reach body temperature.

      Second-order
      Consider putting weights on a mechanical weighing scale. The weight as reported on the measuring dial, will wobble around the correct value at first until reaching equilibrium. An example of this is in clinical practice is the direct measurement of arterial pressure with a transducer. The value of the input fluctuates around a central point.

      Drift is the progressive deterioration of a measurement system’s precision. With time, the measurement deviates from the genuine, calibrated value. The graph between this measurement and the real value should, ideally, be linear (e.g. on the y-axis the measured end-tidal CO2 against true value of the end-tidal CO2). Drift is split into three types: zero-offset, gradient, and zonal drift.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 19 - Bacteria and viruses that are smaller than 0.1 μm in diameter can be...

    Incorrect

    • Bacteria and viruses that are smaller than 0.1 μm in diameter can be filtered out using heat and moisture exchanger (HME) with a typical pore size 0.2 μm. Choose the most appropriate mechanisms of particle capture for most bacteria and viruses.

      Your Answer:

      Correct Answer: Diffusion

      Explanation:

      Warming, humidifying, and filtering inspired anaesthetic gases is done by heat and moisture exchangers (HME) and breathing system filters. They are made of glass fibres materials and are supported by a sturdy frame. Pleating increases the surface area to reduce resistance to air flow and boost efficiency.

      Filters’ effectiveness is determined by the amount and size of particles they keep out of the patient’s airway. The efficiency of filters might be classified as 95, 99.95, or 99.97 percent. Pores with a diameter of 0.2 µm are common. The following are examples of typical particle sizes:
      Red blood cell – 5 µm
      Lymphocyte – 5-8 µm
      Viruses – 0.02-0.3 µm
      Bacteria – 0.5-1 µm
      Depending on particle size, gas flow speed, and charge, particles are collected via a number of processes. Mechanical sieve, interception, diffusion, electrostatic filtration, and inertial impaction are some of the options:

      Sieve:
      The diameter of the particle the filter is supposed to collect is smaller than the apertures of the filter’s fibres.

      Interception:
      When a particle following a gas streamline approaches a fibre within one radius of itself, it becomes attached and captured.
      Diffusion:

      A particle’s random (Brownian) zig-zag path or motion causes it to collide with a fibre.
      By attracting and capturing a particle from within the gas flow, it generates a lower-concentration patch within the gas flow into which another particle diffuses, only to be captured. At low gas velocities and with smaller particles (0.1µm diameter), this is more common.

      Electrostatic:

      These filters use large diameter fibre media and rely on electrostatic charges to improve fine particle removal effectiveness.

      Impaction due to inertia:

      When a particle is too large to respond fast to abrupt changes in streamline direction near a filter fibre, this happens. Because of its inertia, the particle will continue on its original course and collide with the filter fibre. When high gas velocities and dense fibre packing of the filter media are present, this sort of filtration mechanism is most prevalent.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 20 - Among the following, which statement is true regarding electrical safety in an operation...

    Incorrect

    • Among the following, which statement is true regarding electrical safety in an operation theatre?

      Your Answer:

      Correct Answer: The higher the frequency of the current the less risk to the patient

      Explanation:

      The operating theatre is an unusual place with several applications of electrical equipment to the human body. This can lead to potential dangers associated with it that need to be prevented. Electrical safety in the operation theatre is the understanding of how these potential dangers can occur and how they can be prevented.

      Electricity can cause morbidity or mortality by one of the following ways:
      (i) electrocution
      (ii) burns
      (iii) ignition of a flammable material, causing a fire or explosion.

      Electrocution is dependant on factors like duration of contact with electric current, the current pathway and the frequency and size of current.

      Option A: The higher the frequency, the less effects of electrocution on the body.

      Option B & D: Equipment can be classified in classes and types.
      The class designation describes the method used for protection against electrocution. Class I is basic protection, class II is double insulation and class III is safety extra low voltage.
      The type designation describes the degree of protection based on the maximum permissible leakage currents under normal and fault conditions.
      Type B:
      can be class I, II or III but the maximum leakage current must not exceed 100 µA. It is therefore not suitable for direct connection to the heart.
      Type BF
      Similar to type B, but uses an isolated (or floating) circuit.
      Type CF
      Only type CF protect against microshock as they allow leakage currents of 0.05 mA per electrode for class I and 0.01 mA for class II. Microshock is a small leakage current that can cause harm because of direct connection to the heart via transvenous lines or wires, bypassing the impedance of the skin, leading to ventricular fibrillation. Microshock current of 100 μA is sufficient to cause VF.

      Option C: A 75mA electrocution can cause ventricular fibrillation. Use the following as a general guide to understand the effect of current size on the body.
      1 mA – tingling pain
      5 mA – pain
      15 mA – tonic muscular contraction
      50 mA – respiratory muscle paralysis
      75 mA – ventricular fibrillation.

      Option E: Wet skin reduces the resistance to current flow and therefore increases the effects of electrocution.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 21 - Which is the most appropriate statement describing the function of flowmeters? ...

    Incorrect

    • Which is the most appropriate statement describing the function of flowmeters?

      Your Answer:

      Correct Answer: Constant pressure, variable orifice - Heidbrink flowmeters

      Explanation:

      There are different models of flowmeters determined by the applied pressure and its orifice. For instance, the watersight flowmeter functions through applying variable pressure, and it has a variable orifice. In contrast, the bubble flowmeter is operated using a constant pressure and orifice. Flowmeters such as rotameters, Heidbrink and Peak have a constant pressure but variable orifice. On the other hand, flowmeters including a simple pressure gauge, water depression, and pneumotachograph have a constant orifice but variable pressure.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 22 - For a rapid sequence induction of anaesthesia, you are pre-oxygenating a patient using...

    Incorrect

    • For a rapid sequence induction of anaesthesia, you are pre-oxygenating a patient using 100% oxygen and a fresh gas flow equal to the patient's minute ventilation. Which would be the most suitable choice of anaesthetic breathing system in this situation?

      Your Answer:

      Correct Answer: Mapleson A system

      Explanation:

      The Mapleson A (Magill) and coaxial version of the Mapleson A system (Lack circuit) are more efficient for spontaneous breathing than any of the other Mapleson circuits. The fresh gas flow (FGF) required to prevent rebreathing is slightly greater than the alveolar minute ventilation (4-5 litres/minute). This is delivered to the patient through the outer coaxial tube and exhaust gases are moved to the scavenging system through the inner tube. In the Lack circuit, the expiratory valve is located close to the common gas outlet away from the patient end. This is the main advantage of the Lack circuit over the Mapleson A circuit.

      The Mapleson E circuit is a modification of the Ayres T piece and the FGF required to prevent rebreathing is 1.5-2 times the patient’s minute volume.

      The Bain circuit is the coaxial version of the Mapleson D circuit.

      The FGF for spontaneous respiration to avoid rebreathing is 160-200 ml/kg/minute.

      The FGF for controlled ventilation to avoid rebreathing is 70-100 ml/kg/min.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 23 - Which statement most accurately describes the action of increasing the gain in ultrasound...

    Incorrect

    • Which statement most accurately describes the action of increasing the gain in ultrasound imaging?

      Your Answer:

      Correct Answer: Amplifies the returning signal

      Explanation:

      A higher frequency ultrasound comes with a better resolution of the digital image. Ultrasound with a frequency of 15 MHz is best used in imaging of superficial organs such as the thyroid gland, muscles, tendons and breasts whereas deep organs are better imaged using a lower frequency of 2-7MHz because of its ability for deeper penetration but lower resolution. These low frequency probes are also used to diagnose ascites, pleural effusions or can be used in echocardiography.

      The US probe emits and then absorbs a reflected wave. Similar to brightness control, increasing the gain will amplify the return signal which is then attenuated by the tissue. This increases the signal to noise ratio.
      A high frame rate, which basically means the number of times an image is updated onto the screen per second, improves the resolution of a moving 3D image which has become more accurate as the computing power has increased.

      Widening of the image field can be obtained by altering the penetration depth which is obtained by changing the frequency of the US beam.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 24 - The equipment used for patient monitoring in theatre and intensive care settings have...

    Incorrect

    • The equipment used for patient monitoring in theatre and intensive care settings have electrical safety requirements for the protection of hospital staff and patients. Of the different classes of electrical equipment listed, which is least likely to cause a patient to suffer a microshock?

      Your Answer:

      Correct Answer: II (CF)

      Explanation:

      Microshock refers to ventricular fibrillation caused by miniscule amounts of currents or voltages (100-150 microamperes) passing through the myocardial tissue from external cables arising from electrical components within the cardiac muscle, for example, pacemaker electrodes or saline filled venous catheters.
      The risk of shock changes with the construction of electrical equipment in question. The main classes of electrical equipment include: I: Appliances have a protective earth connected to an outer casing which prevents live elements from coming in contact with conductive elements. A fault in this equipment class will result in live elements coming in contact with the outer casing and allowing electrical flow into the protective earth. This triggers the protective fuse to disconnect the electric supply to the appliance.
      II: These appliances have reinforced insulation. In the event of a fault which causes the first layer of insulation to fail, the second layer is able to prevent contact of live elements with outer casing.
      III: These appliances have no insulation to provide safety, and rely solely on the use of separated extra low voltage source (SELV) which limits voltage to 25V AC or 60V DC allowing for a person to come in contact with it without risk of a shock under normal dry conditions. Under wet conditions, voltage supply should be lowered to reduce risk of shock. These devices have no risk of macroshocks, but some risk of microshocks.
      Class I and II electrical appliances are further divided into subtypes developed to limit current leakage in the event of a singular fault:
      B (body): Upper limit of current leakage is 500 µA. This current can cause skin tingling and microshocks, but is not sufficient to cause injury.
      BF (body floating): These appliances have an isolating capacitor or transformer which separate the secondary circuit from the protective earth. The upper limit of current leakage is the same as type B.
      CF (cardiac floating): Upper limit of leakage current during a singular fault is 50 microamps. It is least likely to result in a microshock.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 25 - Which of the following is true about the pulse oximeter? ...

    Incorrect

    • Which of the following is true about the pulse oximeter?

      Your Answer:

      Correct Answer: Is accurate to within +/- 2% in the range of 70% to 100% saturation

      Explanation:

      The pulse oximeter provides a continuous non-invasive measurement of the arterial oxygen saturation. The light emitting diodes (LEDs) produce beams of red and infrared light at 660 nm and 940 nm respectively (not 640 and 960 nm), which travel through a finger (toe, ear lobe or nose) and are then detected by a sensitive photodetector.

      The light absorbed by non-pulsatile tissues is constant (DC), and the non-constant absorption (AC) is the result of arterial blood pulsation. The DC and AC components at 660 and 940 nm are then analysed by the microprocessor and the result is related to the arterial saturation.

      An isosbestic point is a point at which two substances absorb a wavelength of light to the same degree. In pulse oximetry the different absorption profiles of oxyhaemoglobin and deoxyhaemoglobin are used to quantify the haemoglobin saturation (in %). Isosbestic points occur at 590 and 805 nm (not 490 and 805 nm), where the light absorbed is independent of the degree of saturation, and are used as reference points.

      The pulse oximeter is accurate to within +/- 2% in the range of 70% to 100% saturation, and below 70% the readings are extrapolated. Pulse oximeters average their readings every 10 to 20 seconds and thus they cannot detect acute desaturation events. Consequently, they are often referred to as ‘lag’ monitors, due to the time delay in identifying the desaturation episode.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 26 - Which of the following is true regarding a laryngoscope? ...

    Incorrect

    • Which of the following is true regarding a laryngoscope?

      Your Answer:

      Correct Answer: The Wisconsin and Seward are examples of straight blade laryngoscopes

      Explanation:

      Direct laryngoscopy are performed using laryngoscopes and they can be classed according to the shape of the blade as curved or straight.

      Miller, Soper, Wisconsin and Seward are examples of straight blade laryngoscopes. Straight blades are commonly used for intubating neonates and infants but can be used in adults too.

      The tip of the miller blade is advanced over the epiglottis to the tracheal entrance then lifted in order to view the vocal cords.

      The RIGHT-SIDED Macintosh blade is used in adults while the left-sided blade may be used in conditions that make intubation with standard blade difficult e.g. facial deformities.

      The McCoy laryngoscope is based on the STANDARD MACINTOSH blade not Robertshaw’s. It has a lever operated hinged tip, which improves the view during laryngoscopy.

      Polio blade is mounted at an angle of 120-135 degrees to the handle. Originally designed for use during the polio epidemic €‹in intubation patients within iron lung ventilators, it is now useful in patients with conditions like breast hypertrophy, barrel chest, and restricted neck mobility.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 27 - A current flows through a simple electric circuit. Which of the following electrical...

    Incorrect

    • A current flows through a simple electric circuit. Which of the following electrical component configurations has the greatest potential difference?

      Your Answer:

      Correct Answer: Two 5 ohm resistors in series with a passing current of 10 ampere

      Explanation:

      According to the Ohm’s law, the potential difference is defined as

      V(potential difference) = I(current) x (R) resistance

      So, for one resistor of 5 ohms, a 10 ampere current will generate:

      V = I x R
      V = 10 x 5
      V = 50 volts

      The formula for resistors in series can be defined as:

      R(total) = R1+R2

      Hence, when a current of 10 amperes passes through two 5 ohms resistors that are connected in series, the potential difference is:

      V = I x (R1+R2)
      V = 10 x (5+5)
      V = 10 x 10
      V = 100 volts

      The formula for resistors that are connected in a parallel circuit is:

      1/ Rtotal = 1/R1 + 1/R2

      Hence, when a current of 10 amperes passes through two 5 Ω resistors that are connected in a parallel circuit, the potential difference is:

      Rtotal = R1 × R2/ R1 + R2
      Rtotal = 25/10
      Rtotal = 2.5
      V = I x R
      V = 10 x (R1xR2 / R1 + R2)
      V = 10 x (25/10)
      V = 10 x 2.5
      V = 25 volts

      Capacitors are electronic components that have the ability to store energy and charge (Q). The derived SI unit of capacitance (C) is the farad (F), which is equivalent to one coulomb per volt (V). The typical capacitors usually have a very small capacitance range, that ranges from pico to microfarads. On the contrary, supercapacitors can have a capacitance of up to 1-5000 F.

      There are a number of factors that eventually determine the capacitance (C). They are as follows:

      – Larger plate area (A)
      – Closer plate spacing (d)
      – Permittivity (ε) of the material (dielectric) between the plates (vacuum<<<glass), – C = ε × A/d

      The units of stored charge are coulombs (Q), which is equal to the pathway of one ampere of current per second.
      Stored charge, capacitance and voltage can be defined by the following equation:

      V (potential difference across capacitor) = Q(charge) / C (capacitance)

      In a parallel circuit, the formula of capacitors is:

      Ctotal = C1 + C2

      Hence, two 5 farad capacitors in a parallel circuit arrangement with a charge of 100 coulomb and capacitance 10 F will give a potential difference of::

      V = 100/10
      V = 10 volts

      In a series circuit, the formula for capacitors is:

      1/Ctotal = 1/C1 + 1/C2

      Hence, two 5 farad capacitors with a charge of 100 coulomb will give:

      Ctotal = C1 × C2/C1 + C2

      In the example total capacitance = 25/10 = 2.5 F

      V = 100/2.5
      V = 40 volts.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 28 - Dinamap is an automated blood pressure monitoring device. Which of these statements best...

    Incorrect

    • Dinamap is an automated blood pressure monitoring device. Which of these statements best fit its properties?

      Your Answer:

      Correct Answer: The cuff should be positioned at the same level as the heart

      Explanation:

      Dinamap continuously measures the systolic, diastolic and mean arterial pressure along with pulse rate, thereby providing a continuous monitoring of the blood pressure using the osscillitonometric principle of measurement.

      The device loses accuracy towards the extremes of BP and is more accurate with systolic compared with diastolic pressure. In arrhythmias such as AF, the devices are also inaccurate due to the major fluctuations associated with the individual pulse pressure variations.

      The manual BP device is still the gold standard for BP measurement and monitoring.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 29 - With respect to the peripheral nerve stimulators, which one is used to perform...

    Incorrect

    • With respect to the peripheral nerve stimulators, which one is used to perform nerve blocking?

      Your Answer:

      Correct Answer: Using an insulated needle improves the success rate for the block

      Explanation:

      The nerve stimulators deliver a stimulus lasting for 1-2 milliseconds (not second) to perform nerve blockage.

      There are just 2 leads (not 3); one for the skin and other for the needle.

      Prior to the administration of the local anaesthesia, a current of 0.25 – 0.5 mA (not 1-2mA) at the frequency of 1-2 Hz is preferred.

      If the needle tip is close to the nerve, muscular contraction could be possible at the lowest possible current.

      Insulated needles have improved the block success rate, as the current is only conducting through needle tip.

      Stimulus to the femoral nerve which is placed in the mid lingual line causes withdrawer of the quadriceps and knee extension, that’s the dancing patella ( not plantar flexion).

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 30 - From the following electromagnetic waves, which one has the shortest wavelength? ...

    Incorrect

    • From the following electromagnetic waves, which one has the shortest wavelength?

      Your Answer:

      Correct Answer: X rays

      Explanation:

      Electromagnetic waves are categorized according to their frequency or equivalently according to their wavelength. Visible light makes up a small part of the full electromagnetic spectrum.

      Electromagnetic waves with shorter wavelengths and higher frequencies include ultraviolet light, X-rays, and gamma rays. Electromagnetic waves with longer wavelengths and lower frequencies include infrared light, microwaves, and radio and televisions waves.

      Different electromagnetic waves according to their wavelength from shorter to longer are X-rays, ultraviolet radiations, visible light, infrared radiation, radio waves. X-ray among electromagnetic waves has the shortest wavelength and higher frequency with wavelengths ranging from 10*-8 to 10* -12 and corresponding frequencies.

    • This question is part of the following fields: