  • Question 1 - A measuring system's response to change is complex, yet it can be mathematically modelled. Which of the following terms best characterises a pressure transducer's responsiveness to blood pressure changes? ...

    Incorrect

    • A measuring system's response to change is complex, yet it can be mathematically modelled. Which of the following terms best characterises a pressure transducer's responsiveness to blood pressure changes?

      Your Answer: Static response

      Correct Answer: Dynamic second-order response

      Explanation:

      The static-response defines how a measuring system behaves while it is in equilibrium (i.e. when the measured values are not changing). If the value being measured changes over time, the reaction of a measuring system will change as well which would be a dynamic response.
      The dynamic response of a measuring system can be subdivided into zero-order, first-order and second-order responses:

      Zero-order:
      Consider a thermometer that has been left in a room for a week. The thermometer will display the current ambient temperature when you enter the room.

      First-order:
      Consider the use of a mercury thermometer to check a patient’s temperature. It is comprised of a mercury column that expands as it warms up. The scale’s initial temperature is room temperature, but when it’s placed under the patient’s tongue, the temperature readings rise until they reach body temperature.

      Second-order
      Consider putting weights on a mechanical weighing scale. The weight as reported on the measuring dial, will wobble around the correct value at first until reaching equilibrium. An example of this is in clinical practice is the direct measurement of arterial pressure with a transducer. The value of the input fluctuates around a central point.

      Drift is the progressive deterioration of a measurement system’s precision. With time, the measurement deviates from the genuine, calibrated value. The graph between this measurement and the real value should, ideally, be linear (e.g. on the y-axis the measured end-tidal CO2 against true value of the end-tidal CO2). Drift is split into three types: zero-offset, gradient, and zonal drift.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      12.4
      Seconds
  • Question 2 - Health workers are at increased risks of anaesthetic exposure. Therefore, The Control of...

    Incorrect

    • Health workers are at increased risks of anaesthetic exposure. Therefore, The Control of Substances Hazardous to Health (COSHH) regulations require measures implemented to assess and control the risks related to this exposure. Among the following control measures, which one is most likely to limit potentially harmful exposure to anaesthetic agents within an operating theatre?

      Your Answer: Active scavenging systems

      Correct Answer: Total intravenous anaesthesia (TIVA)

      Explanation:

      Exposure to anaesthetic hazards is one among the occupational exposures in manipulating toxic agents or inhaling toxic gases during anaesthetic practices.

      Toxic gases mainly nitrous oxide, is one of the most gaseous anaesthetic agents that constitutes an important source of pollution. One of the safe and effective technics used in anaesthesia and reducing the amount of pollution is the Total Intravenous Anaesthesia (TIVA) which consists of using opioids in analgesia and propofol for the induction and the maintenance of anaesthesia. It refers to the administration intravenously of an anaesthetic, sedative, and/or tranquilizer. A less polluting but not the best way to get rid of the toxic aesthetic agents is the scavenger system that collects and expels the gas outside the medical environment. Yet, this technique still represents a hazard for the environment and still increase the risk of exposure for the health workers and clinical staff.

      Fume cupboards are also not recommended to use because of their high pollution potency, mainly of the air resulting in a great harm for medical workers.

      Supraglottic airways as well as the Air Changes per Hour technics could be harmful for both patients and health workers, increasing the risks of transmitted diseases, namely nosocomial infections.

      Therefore, the Total Intravenous Anaesthesia technique (TIVA) is most likely to be safe and recommended to use.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      4.6
      Seconds
  • Question 3 - With respect to the peripheral nerve stimulators, which one is used to perform...

    Incorrect

    • With respect to the peripheral nerve stimulators, which one is used to perform nerve blocking?

      Your Answer:

      Correct Answer: Using an insulated needle improves the success rate for the block

      Explanation:

      The nerve stimulators deliver a stimulus lasting for 1-2 milliseconds (not second) to perform nerve blockage.

      There are just 2 leads (not 3); one for the skin and other for the needle.

      Prior to the administration of the local anaesthesia, a current of 0.25 – 0.5 mA (not 1-2mA) at the frequency of 1-2 Hz is preferred.

      If the needle tip is close to the nerve, muscular contraction could be possible at the lowest possible current.

      Insulated needles have improved the block success rate, as the current is only conducting through needle tip.

      Stimulus to the femoral nerve which is placed in the mid lingual line causes withdrawer of the quadriceps and knee extension, that’s the dancing patella ( not plantar flexion).

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 4 - A 74-year-old with a VVI pacemaker is undergoing a hip replacement. Which of...

    Incorrect

    • A 74-year-old with a VVI pacemaker is undergoing a hip replacement. Which of the following is most likely to predispose him to an electrical hazard?

      Your Answer:

      Correct Answer: Use of cutting unipolar diathermy

      Explanation:

      A single chamber pacemaker was implanted in the patient. In VVI mode, a pacemaker paces and senses the ventricle while being inhibited by a perceived ventricular event. The most likely electrical hazard from diathermy is electromagnetic interference (EMI).

      EMI has the potential to cause the following: Inhibition of pacing
      Asynchronous pacing
      Reset to backup mode
      Myocardial burns, and
      Trigger VF.

      Diathermy entails the implementation of high-frequency electrical currents to produce heat and either make incisions or induce coagulation. Monopolar cautery involves disposable cautery pencils and electrosurgical diathermy units. In typical monopolar cautery, an electrical plate is placed on the patient’s skin and acts as an electrode, while the current passes between the instrument and the plate. Monopolar diathermy can therefore interfere with implanted metal devices and pacemaker function.

      Bipolar diathermy, where the current passes between the forceps tips and not through the patient and is less likely to generate EMI.

      Whilst the presence of a CVP line may in theory predispose the patient to microshock, the use of prerequisite CF electrical equipment makes this very unlikely. The presence of a CVP line and pacemaker does not therefore unduly increase the risk of an electrical hazard.

      Isolating transformers are used to protect secondary circuits and individuals from electrical shocks. There is no step-up or step-down voltage (i.e. there is a ratio of 1 to 1 between the primary and secondary windings).

      A ground (or earth) wire is normally connected to the metal case of an operating table to protect patients from accidental electrocution. In the event that a fault allows a live wire to make contact with the metal table (broken cable, loose connection etc.) it becomes live. The earth will provide an immediate path for current to safely flow through and so the table remains safe to touch. Being a low resistance path, the earth lets a large current flow through it when the fault occurs ensuring that the fuse or RCD will quickly blow. Without an operating table earth, the patient is not at more risk of an electrical hazard because of the pacemaker.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 5 - A current flows through a simple electric circuit. Which of the following electrical...

    Incorrect

    • A current flows through a simple electric circuit. Which of the following electrical component configurations has the greatest potential difference?

      Your Answer:

      Correct Answer: Two 5 ohm resistors in series with a passing current of 10 ampere

      Explanation:

      According to the Ohm’s law, the potential difference is defined as

      V(potential difference) = I(current) x (R) resistance

      So, for one resistor of 5 ohms, a 10 ampere current will generate:

      V = I x R
      V = 10 x 5
      V = 50 volts

      The formula for resistors in series can be defined as:

      R(total) = R1+R2

      Hence, when a current of 10 amperes passes through two 5 ohms resistors that are connected in series, the potential difference is:

      V = I x (R1+R2)
      V = 10 x (5+5)
      V = 10 x 10
      V = 100 volts

      The formula for resistors that are connected in a parallel circuit is:

      1/ Rtotal = 1/R1 + 1/R2

      Hence, when a current of 10 amperes passes through two 5 Ω resistors that are connected in a parallel circuit, the potential difference is:

      Rtotal = R1 × R2/ R1 + R2
      Rtotal = 25/10
      Rtotal = 2.5
      V = I x R
      V = 10 x (R1xR2 / R1 + R2)
      V = 10 x (25/10)
      V = 10 x 2.5
      V = 25 volts

      Capacitors are electronic components that have the ability to store energy and charge (Q). The derived SI unit of capacitance (C) is the farad (F), which is equivalent to one coulomb per volt (V). The typical capacitors usually have a very small capacitance range, that ranges from pico to microfarads. On the contrary, supercapacitors can have a capacitance of up to 1-5000 F.

      There are a number of factors that eventually determine the capacitance (C). They are as follows:

      – Larger plate area (A)
      – Closer plate spacing (d)
      – Permittivity (Δ) of the material (dielectric) between the plates (vacuum<<<glass), – C = Δ × A/d

      The units of stored charge are coulombs (Q), which is equal to the pathway of one ampere of current per second.
      Stored charge, capacitance and voltage can be defined by the following equation:

      V (potential difference across capacitor) = Q(charge) / C (capacitance)

      In a parallel circuit, the formula of capacitors is:

      Ctotal = C1 + C2

      Hence, two 5 farad capacitors in a parallel circuit arrangement with a charge of 100 coulomb and capacitance 10 F will give a potential difference of::

      V = 100/10
      V = 10 volts

      In a series circuit, the formula for capacitors is:

      1/Ctotal = 1/C1 + 1/C2

      Hence, two 5 farad capacitors with a charge of 100 coulomb will give:

      Ctotal = C1 × C2/C1 + C2

      In the example total capacitance = 25/10 = 2.5 F

      V = 100/2.5
      V = 40 volts.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 6 - What feature is found in the flowmeters of modern anaesthetic machines ? ...

    Incorrect

    • What feature is found in the flowmeters of modern anaesthetic machines ?

      Your Answer:

      Correct Answer: The use of stannic oxide increases the accuracy of flowmeters

      Explanation:

      Flowmeters measure the rate at which a specific gas, that the flowmeter has been calibrated for, passes through. This calibration is done at room temperature and standard atmospheric pressure with an accuracy of +/- 2%.

      Reading the flowmeter is done from the top of a bobbin (the midpoint of a ball). Oxygen is the last gas to be added downstream to the mixture delivered to the back bar as a safety feature. This prevents delivery of a hypoxic mixture.

      Inaccurate flow measurements occur when the bobbin sticks to the inside wall of the flowmeter. Stannic oxide has been used as a successful antistatic substance thus, reducing the aforementioned risk.

      Carbon dioxide being easily delivered is found on some older machines, but those attached flowmeters are limited by a maximum flow of 500 ml /min. Thus avoiding the delivery of a hypercarbic mixture.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 7 - Regarding laminar gas flow, which of the following options has the most influence...

    Incorrect

    • Regarding laminar gas flow, which of the following options has the most influence on laminar flow?

      Your Answer:

      Correct Answer: Diameter of tube

      Explanation:

      Laminar flow can be defined as the motion of a fluid where every particle in the fluid follows the same path of its previous particles. The following are properties of laminar flow of gas or fluids:

      1. Smooth unobstructed flow of gas through a tube of relatively uniform diameter
      2. Few directional changes
      3. Slow, steady flow through straight smooth, rigid, large calibre, cylindrical tube
      4. Outer layer flow slower than the centre due to friction, results in discrete cylindrical layers, or streamlines
      5. Double flow by doubling pressure as long as the flow pattern remains laminar

      Poiseuille’s Law relates the factors that determine laminar flow. It indicates the degree of resistance to fluid flow through a tube. The resistance to fluid flow through a tube is directly related to the length, flow and viscosity; and inversely related to the radius of the tube to the fourth power. This means that, when the radius is doubled, there is increase in flow by a factor of 16.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 8 - A laser is a device that stimulates atoms or molecules to emit light...

    Incorrect

    • A laser is a device that stimulates atoms or molecules to emit light at particular wavelengths and amplifies that light, typically producing a very narrow beam of radiation. This can be of visible, infrared, or ultraviolet wavelengths. They have been widely utilized in theatre environment. Which of the following safety measures is most likely to reduce chances of eye injury to the theatre personnel?

      Your Answer:

      Correct Answer: Wearing laser protective goggles

      Explanation:

      Eye damage is the most common potential hazard associated with laser energy. Everyone in the laser treatment room has the risk of eye exposure when working with a Class 3b or Class 4 healthcare laser system, and damage to various structures in the eye depending on wavelength of the laser if they are unprotected.

      Red and near-infrared light (400-1400 nm) has very high penetration power. The light causes painless burns on the retina after it is absorbed by melanin in the pigment epithelium just behind the photoreceptors.

      Infrared radiation (IR), or infrared light (>1060 nm), is a type of radiant energy that’s invisible to human eyes and hence won’t elicit the protective blink.

      Ultraviolet light (<400 nm) is also a form of electromagnetic radiation which is can penetrate the cornea and be absorbed by the iris or the pupil and cause burn injuries or cataract occur due to irreversible photochemical retinal damage.

      Safety eyewear is the best method of providing eye protection and are designed to absorb light specific to the laser being used. Laser protective eyewear (LPE) includes glasses or goggles of proper optical density (OD). The lenses should not be glass or plastic. The LPE should withstand direct and diffuse scattered laser beams.

      The laser protection supervisor (LPS) or LSO is an individual who is responsible for any clinical area in which lasers are used. They are expected to have a certain level of equipment and determine what control measures are appropriate, for each individual system, but their presence does not guarantee the chances of having an eye injury.

      Class 1 lasers are generally safe under every conceivable condition and is not likely to cause any eye damage. Class 3b or Class 4 medical laser systems are utilized in healthcare which have their own safety precautions.

      Polarized spectacles can make your eyes more comfortable by eliminated glare, however, they will not be able to offer any protection against wavelengths at which laser act.
      Using short bursts to reduce energy is also not correct as it would still be harmful to eye.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 9 - A transport ventilator is powered by an air/oxygen mix using a full oxygen...

    Incorrect

    • A transport ventilator is powered by an air/oxygen mix using a full oxygen cylinder (class CD) with an internal capacity of 2 litres, and pressure of 23,000 kPa, with a gas flow of 4 litres/minute. The ventilator also has a control resulting in an additional gas consumption of 1 litre/minute. How long will it take for the cylinder to empty?

      Your Answer:

      Correct Answer: 92 minutes

      Explanation:

      The Drager OxylogÂź 1000 is a pneumatically powered, time-dependent, volume-titrated emergency ventilator with a pressure limit. It is compatible with CD cylinder oxygen. The CD cylinder is a strong and lightweight cylinder usually composed of aluminium or Kevlar. The internal cylinder volume is 2 litres, and the pressure of a full cylinder is 230 bar. The volume of the full cylinder is determined by applying Boyle’s law: P1 × V1 = P2 × V2

      Where:
      P1= pressure of a full cylinder (230 bar)
      V1= volume of oxygen at that pressure (2 litres)
      P2= final pressure (1 bar), and
      V2= volume of oxygen in the full cylinder.

      Substituting values into the equation:

      230 × 2 = 1 x V2
      V2 = 460 litres. The flow of fresh gas is 4 litres/minute + 1 litre/minute required by the control, making a total of 5 litres/minute. The amount of time it takes for the cylinder to empty would be the total volume of oxygen in the full cylinder divided by the amount of oxygen expelled per minute: 460/5 = 92, meaning it would take 92 minutes for the cylinder to empty.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 10 - A sevoflurane vaporiser with a 2 percent setting and a 200 kPa ambient...

    Incorrect

    • A sevoflurane vaporiser with a 2 percent setting and a 200 kPa ambient pressure is used. At this pressure, which of the following options best represents vaporiser output?

      Your Answer:

      Correct Answer: The output is 1% because the saturated pressure of sevoflurane is unaffected by ambient pressure

      Explanation:

      Ambient pressure has no effect on a volatile agent’s saturated vapour pressure (SVP). At a temperature of 20°C, the SVP of sevoflurane is approximately 21 kPa, or 21% of atmospheric pressure (100 kPa).

      The SVP of sevoflurane remains the same when the ambient pressure is doubled to 200 kPa, but the output of the vaporiser is halved, now 21 percent of 200 kPa, equalling 10.5 percent. The vaporiser’s output has increased to 1%, but the partial pressure output has remained unchanged. The splitting ratio will not change because it is determined by temperature changes.

      Calculations can be made as follows:

      Vaporizer output % (ambient pressure) = % volatile (calibrated) x 100 kPa calibrated pressure/ambient pressure
      2% = 2% (dialled) × 100/100
      2% of 100 = 2 kPa

      Altitude, pressure 50 kPa
      4% = 2% (dialled) × 100/50
      4% of 50 = 2 kPa

      High pressure at 200 kPa
      1% = 2% (dialled) × 100/200
      1% of 200 = 2 kPa

      Sevoflurane has a boiling point of 58°C and, unlike desflurane (which has a boiling point of 22.8°C), does not need to be heated and pressurised with a Tec 6 vaporiser.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 11 - When the volume of gas is measured, its value should be supported by...

    Incorrect

    • When the volume of gas is measured, its value should be supported by a suitable abbreviation. From the following options, what does the STPD stand for?

      Your Answer:

      Correct Answer: Standard Temperature and Pressure, Dry

      Explanation:

      Gas is composed of large numbers of molecules moving in random directions, separated by distances. They undergo perfectly elastic collisions with each other and the walls of a container and transfer kinetic energy in form of heat. These assumptions bring the characteristics of gases within the range and reasonable approximation to a real gas, particularly how any change in temperature and pressure affect the behaviour of gas. According to different theories and laws proposed, mathematical equations are derived to calculate the volume of gas, also different abbreviations are being used according to given conditions. The abbreviations used are ATP, BTPS, and STPD.
      ATP stands for ambient temperature and barometric pressure, it is used to describe the conditions under which volume of gas is measured.
      BTPS stands for body temperature and pressure saturated with water vapor. These are conditions under which volume of gas exist and all results of lung volume determination should be quoted at BTPS.
      STPD stands for standard temperature and pressure, dry (0C and 760 mm Hg). These are the conditions that are used to describe quantities of individual gases exchanged in the lungs.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 12 - Which of the following is true about the pulse oximeter? ...

    Incorrect

    • Which of the following is true about the pulse oximeter?

      Your Answer:

      Correct Answer: Is accurate to within +/- 2% in the range of 70% to 100% saturation

      Explanation:

      The pulse oximeter provides a continuous non-invasive measurement of the arterial oxygen saturation. The light emitting diodes (LEDs) produce beams of red and infrared light at 660 nm and 940 nm respectively (not 640 and 960 nm), which travel through a finger (toe, ear lobe or nose) and are then detected by a sensitive photodetector.

      The light absorbed by non-pulsatile tissues is constant (DC), and the non-constant absorption (AC) is the result of arterial blood pulsation. The DC and AC components at 660 and 940 nm are then analysed by the microprocessor and the result is related to the arterial saturation.

      An isosbestic point is a point at which two substances absorb a wavelength of light to the same degree. In pulse oximetry the different absorption profiles of oxyhaemoglobin and deoxyhaemoglobin are used to quantify the haemoglobin saturation (in %). Isosbestic points occur at 590 and 805 nm (not 490 and 805 nm), where the light absorbed is independent of the degree of saturation, and are used as reference points.

      The pulse oximeter is accurate to within +/- 2% in the range of 70% to 100% saturation, and below 70% the readings are extrapolated. Pulse oximeters average their readings every 10 to 20 seconds and thus they cannot detect acute desaturation events. Consequently, they are often referred to as ‘lag’ monitors, due to the time delay in identifying the desaturation episode.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 13 - Which statement most accurately describes the action of increasing the gain in ultrasound...

    Incorrect

    • Which statement most accurately describes the action of increasing the gain in ultrasound imaging?

      Your Answer:

      Correct Answer: Amplifies the returning signal

      Explanation:

      A higher frequency ultrasound comes with a better resolution of the digital image. Ultrasound with a frequency of 15 MHz is best used in imaging of superficial organs such as the thyroid gland, muscles, tendons and breasts whereas deep organs are better imaged using a lower frequency of 2-7MHz because of its ability for deeper penetration but lower resolution. These low frequency probes are also used to diagnose ascites, pleural effusions or can be used in echocardiography.

      The US probe emits and then absorbs a reflected wave. Similar to brightness control, increasing the gain will amplify the return signal which is then attenuated by the tissue. This increases the signal to noise ratio.
      A high frame rate, which basically means the number of times an image is updated onto the screen per second, improves the resolution of a moving 3D image which has become more accurate as the computing power has increased.

      Widening of the image field can be obtained by altering the penetration depth which is obtained by changing the frequency of the US beam.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 14 - When compared to unipolar diathermy, which of the following is more specific to...

    Incorrect

    • When compared to unipolar diathermy, which of the following is more specific to bipolar diathermy?

      Your Answer:

      Correct Answer: Has a power output of up to 140 joules per second

      Explanation:

      Electrocautery, also known as diathermy, is a technique for coagulation, tissue cutting, and fulguration that uses a high-frequency current to generate heat (cell destruction from dehydration).

      The two electrodes in bipolar diathermy are the tips of forceps, and current passes between the tips rather than through the patient. Bipolar diathermy’s power output (40-140 W) is lower than unipolar diathermy’s typical output (400 W). There is no earthing in the bipolar circuit.

      A cutting electrode and a indifferent electrode in the form of a metal plate are used in unipolar diathermy. The high-frequency current completes a circuit by passing through the patient from the active electrode to the metal plate. When used correctly, the current density at the indifferent electrode is low, and the patient is unlikely to be burned. Between the patient plate and the earth is placed an isolating capacitor. This has a low impedance to a high frequency current, such as diathermy current, and is used in modern diathermy machines. The capacitor has a high impedance to current at 50 Hz, which protects the patient from electrical shock.

      High frequency currents (500 KHz – 1 MHz) are used in both unipolar and bipolar diathermy, which can cause tissue damage and interfere with pacemaker function (less so with bipolar diathermy).

      The effect of diathermy is determined by the current density and waveform employed. The current is a pulsed square wave pattern in coagulation mode and a continuous square wave pattern in cutting mode.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 15 - Which of the following statements is TRUE regarding an epidural set? ...

    Incorrect

    • Which of the following statements is TRUE regarding an epidural set?

      Your Answer:

      Correct Answer: 19G Tuohy needles have 0.5 cm markings

      Explanation:

      A paediatric 19G Tuohy catheter is available that is 5cm in length and has 0.5cm markings

      18G Tuohy catheters are generally 9 to 10cm to hub

      Distal end of catheter is angled (15 to 30 degrees) and closed to avoid puncturing the dura

      Epidural mesh are usually 0.2 microns and are used to filter bacteria and viruses to ensure sterility of procedure

      Transparent catheters are 90cm long with diameters depending on gauge size. It has 1cm graduations from 5 to 20cm to ensure they have been inserted amply and removed completely. Distal end is smooth which can be open or closed (with lateral openings).

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 16 - Regarding the emergency oxygen flush, which is true? ...

    Incorrect

    • Regarding the emergency oxygen flush, which is true?

      Your Answer:

      Correct Answer: May lead to awareness if used inappropriately

      Explanation:

      When the emergency oxygen flush is pressed, 100% oxygen is supplied from the common gas outlet. This gas bypasses BOTH flowmeters and vaporisers. The flow of oxygen is usually 45 l/min at a PRESSURE OF 400 kPa.

      There is an increased risk of pulmonary barotrauma when the emergency flush is pressed, especially when anaesthetising paediatric patients.

      The inappropriate use of the flush causes dilution of anaesthetic gases and this increases the possibility of anaesthetic awareness.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 17 - The Control of Substances Hazardous to Health (COSHH) regulations recommend air supply rates...

    Incorrect

    • The Control of Substances Hazardous to Health (COSHH) regulations recommend air supply rates to specific environments. Which of the following statements is true?

      Your Answer:

      Correct Answer: Preparation rooms receive a volume of 0.1 m3 of air per second

      Explanation:

      Control of Substances Hazardous to Health (COSHH) was established by government under the Health and Safety at Work act in 1989. Their employers work on identification and management of those substances that are dangerous to health. The implications for anaesthetists include gas scavenging, equipment contamination and environmental safety. Adequate ventilation is required in areas where anaesthetic gases are present. The minimum air supply that is legally required in each specific area is: Operating theatres: 0.65 m3/second. Anaesthetic rooms: 0.15 m3/s. Preparation rooms: 0.1 m3/s. Recovery rooms need 15 air changes per hour.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 18 - Regarding the classification of breathing systems, which of the following is true? ...

    Incorrect

    • Regarding the classification of breathing systems, which of the following is true?

      Your Answer:

      Correct Answer: The Conway classification describes a functional classification based on whether a CO2 absorber is required

      Explanation:

      Breathing system is an assembly of components which connects patient’s airway to anaesthesia machine through which controlled composition of gas mixture is dispensed. It delivers gas to the patient, removes expired gas and controls the temperature and humidity of the inspired mixture. It allows spontaneous, controlled, or assisted respiration. It may also provide ports for gas sampling, airway pressure, flow and volume monitoring.

      Breathing systems have been classified by Conway and Mapleson.
      Conway suggested a functional classification:
      – Circuits requiring a CO2 absorber
      – Circuits not requiring a CO2 absorber

      William Mapleson designated varying arrangements of breathing system components (masks, breathing tubes, fresh gas flow inlets, adjustable pressure-limiting valves, and reservoir bags) as Mapleson A-E circuits.
      Mapleson A: Arranged as FGF inlet, reservoir bag, APL valve, mask.
      In this circuit, because the reservoir bag is between the FGF inlet valve and the APL valve, expired gas from the patient may re-enter the system and fill the reservoir bag during controlled ventilation. This is the most efficient system for spontaneous breathing as the FGF must only be equal to a patient’s minute ventilation to prevent rebreathing.

      Mapleson B: Arranged as reservoir bag, FGF inlet, APL valve, mask.
      In this circuit, the FGF inlet is closer to the APL valve, which helps prevent the rebreathing concern in the Mapleson A circuit as above during controlled ventilation.

      Mapleson C: Arranged as reservoir bag, FGF inlet, APL valve, mask.
      In this circuit, the arrangement is the same as the Mapleson B circuit. However, this circuit is shorter as it does not contain elongated corrugated tubing. This circuit also has the FGF inlet close to the APL valve to aid in preventing rebreathing.

      Mapleson D: Arranged as reservoir bag, APL valve, FGF inlet, and mask.
      In this circuit, the arrangement interchanges the FGF inlet and APL valve of the Mapleson A circuit. This system prevents rebreathing by directing FGF towards the APL valve rather than towards the patient during exhalation.

      Mapleson E: Arranged as corrugated tubing, FGF inlet, and mask.
      In this circuit, there is no reservoir bag and no APL valve. Given the inability to alter the pressure of the circuit, this is ideal for spontaneously ventilating neonates or paediatric patients where low-pressure ventilation is desired. The system prevents rebreathing, similar to the Mapleson D circuit.

      Jackson Rees later modified the Mapleson E by adding an open ended bag, which has since become known as the Mapleson F.
      Mapleson F: Arranged as APL valve directly connected to reservoir bag, corrugated tubing, FGF inlet, and mask.
      The system prevents rebreathing similarly to Mapleson D by directing FGF towards the APL valve.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 19 - Which of the following statements is true about oxygen face masks? ...

    Incorrect

    • Which of the following statements is true about oxygen face masks?

      Your Answer:

      Correct Answer: The oxygen concentration delivered by high air flow oxygen enrichment devices is not dependent on the respiratory pattern of the patient

      Explanation:

      The normal peak inspiratory flow in healthy individuals is 20-30 L/min during each normal tidal ventilation. This is expected to increase with greater respiratory rate and deeper inspiration.

      Face masks are used to facilitate the delivery of oxygen from a breathing system to a patient. Face masks can be divided into two types: fixed performance or variable performance devices.

      In fixed performance devices (also known as high air flow oxygen enrichment or HAFOE), fixed inspired oxygen concentration is delivered to the patent, independent and greater than that of the patient’s peak inspiratory flow rate (PIFR). No random entrainment is expected to occur at the time of PIFR, hence, the oxygen concentration in HAFOE devices is not dependent on the patient’s respiratory pattern.

      Moreover, in HAFOE masks, the concentration of oxygen at a given oxygen flow rate is determined by the size of the constriction; a device with a greater entrainment aperture delivers a lower oxygen concentration. Therefore, a 40% Venturi device will have lesser entrainment aperture when compared to a 31% Venturi. Venturi masks allow relatively fixed concentrations of supplemental oxygen to be inspired e.g. 24%, 28%, 31%, 35%, 40% and 60% oxygen. These are colour coded and marked with the recommended oxygen flow rate.

      Variable performance devices deliver variable inspired oxygen concentration to the patient, and is dependent on the PIFR. The PIFR can often exceed the flow rate at which oxygen or an oxygen/air mixture is supplied by the device, depending on a patient’s inspiratory effort. In addition, these masks allow expired air to be released through the holes in the sides of the mask. Thus, with increased respiratory rate, rebreathing of alveolar gas from inside the mask may occur.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 20 - Bacteria and viruses that are smaller than 0.1 ĂŽÂŒm in diameter can be...

    Incorrect

    • Bacteria and viruses that are smaller than 0.1 ĂŽÂŒm in diameter can be filtered out using heat and moisture exchanger (HME) with a typical pore size 0.2 ĂŽÂŒm. Choose the most appropriate mechanisms of particle capture for most bacteria and viruses.

      Your Answer:

      Correct Answer: Diffusion

      Explanation:

      Warming, humidifying, and filtering inspired anaesthetic gases is done by heat and moisture exchangers (HME) and breathing system filters. They are made of glass fibres materials and are supported by a sturdy frame. Pleating increases the surface area to reduce resistance to air flow and boost efficiency.

      Filters’ effectiveness is determined by the amount and size of particles they keep out of the patient’s airway. The efficiency of filters might be classified as 95, 99.95, or 99.97 percent. Pores with a diameter of 0.2 ”m are common. The following are examples of typical particle sizes:
      Red blood cell – 5 ”m
      Lymphocyte – 5-8 ”m
      Viruses – 0.02-0.3 ”m
      Bacteria – 0.5-1 ”m
      Depending on particle size, gas flow speed, and charge, particles are collected via a number of processes. Mechanical sieve, interception, diffusion, electrostatic filtration, and inertial impaction are some of the options:

      Sieve:
      The diameter of the particle the filter is supposed to collect is smaller than the apertures of the filter’s fibres.

      Interception:
      When a particle following a gas streamline approaches a fibre within one radius of itself, it becomes attached and captured.
      Diffusion:

      A particle’s random (Brownian) zig-zag path or motion causes it to collide with a fibre.
      By attracting and capturing a particle from within the gas flow, it generates a lower-concentration patch within the gas flow into which another particle diffuses, only to be captured. At low gas velocities and with smaller particles (0.1”m diameter), this is more common.

      Electrostatic:

      These filters use large diameter fibre media and rely on electrostatic charges to improve fine particle removal effectiveness.

      Impaction due to inertia:

      When a particle is too large to respond fast to abrupt changes in streamline direction near a filter fibre, this happens. Because of its inertia, the particle will continue on its original course and collide with the filter fibre. When high gas velocities and dense fibre packing of the filter media are present, this sort of filtration mechanism is most prevalent.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 21 - All of the following statements about pH electrode are incorrect except: ...

    Incorrect

    • All of the following statements about pH electrode are incorrect except:

      Your Answer:

      Correct Answer: A semi-permeable membrane reduces protein contamination

      Explanation:

      Pulse oximeters combine the principles of oximetry and plethysmography to noninvasively measure oxygen saturation in arterial blood. A sensor containing two or three light emitting diodes and a photodiode is placed across a perfused body part, commonly a finger, to be transilluminated. Oximetry depends on oxyhaemoglobin and deoxyhaemoglobin, and their ability to absorb the beams of light produced by the light emitting diodes: red light at 660 nm and infrared light at 960 nm.

      The isosbestic point is the point wherein two different substances absorb light to the same extent. For oxyhaemoglobin and deoxyhaemoglobin, the points are at 590 nm and 805 nm. These are considered reference points where light absorption is independent of the degree of saturation.

      Non-constant absorption of light is often due to the presence of an arterial pulsation, whilst constant absorption of light is seen in non-pulsatile tissues.

      Most pulse oximeters are inaccurate at low SpO2, but is accurate at +/- 2% within the range of 70% to 100% SpO2. All pulse oximeters demonstrate a delay in between changes in SaO2 and SpO2, and display average readings every 10 to 20 seconds, hence they are unable to detect acute desaturation episodes.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 22 - Among the following, which statement is true regarding electrical safety in an operation...

    Incorrect

    • Among the following, which statement is true regarding electrical safety in an operation theatre?

      Your Answer:

      Correct Answer: The higher the frequency of the current the less risk to the patient

      Explanation:

      The operating theatre is an unusual place with several applications of electrical equipment to the human body. This can lead to potential dangers associated with it that need to be prevented. Electrical safety in the operation theatre is the understanding of how these potential dangers can occur and how they can be prevented.

      Electricity can cause morbidity or mortality by one of the following ways:
      (i) electrocution
      (ii) burns
      (iii) ignition of a flammable material, causing a fire or explosion.

      Electrocution is dependant on factors like duration of contact with electric current, the current pathway and the frequency and size of current.

      Option A: The higher the frequency, the less effects of electrocution on the body.

      Option B & D: Equipment can be classified in classes and types.
      The class designation describes the method used for protection against electrocution. Class I is basic protection, class II is double insulation and class III is safety extra low voltage.
      The type designation describes the degree of protection based on the maximum permissible leakage currents under normal and fault conditions.
      Type B:
      can be class I, II or III but the maximum leakage current must not exceed 100 ”A. It is therefore not suitable for direct connection to the heart.
      Type BF
      Similar to type B, but uses an isolated (or floating) circuit.
      Type CF
      Only type CF protect against microshock as they allow leakage currents of 0.05 mA per electrode for class I and 0.01 mA for class II. Microshock is a small leakage current that can cause harm because of direct connection to the heart via transvenous lines or wires, bypassing the impedance of the skin, leading to ventricular fibrillation. Microshock current of 100 ĂŽÂŒA is sufficient to cause VF.

      Option C: A 75mA electrocution can cause ventricular fibrillation. Use the following as a general guide to understand the effect of current size on the body.
      1 mA – tingling pain
      5 mA – pain
      15 mA – tonic muscular contraction
      50 mA – respiratory muscle paralysis
      75 mA – ventricular fibrillation.

      Option E: Wet skin reduces the resistance to current flow and therefore increases the effects of electrocution.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 23 - Regarding the carbon dioxide monitoring, which of the following statements is correct? ...

    Incorrect

    • Regarding the carbon dioxide monitoring, which of the following statements is correct?

      Your Answer:

      Correct Answer: Carbon dioxide absorbs infrared radiation at 4.28 ”m

      Explanation:

      Carbon dioxide (CO2), is a carbonic gas made up of two dissimilar atoms, namely one carbon atom and two oxygen atoms. Capnography is a technique used to measure carbon dioxide during a respiratory cycle, and it consists in calculating the concentration of the partial pressure of CO2, through the absorption of the infrared light, namely that CO2 absorbs infrared radiation at a wavelength of 4.28 ”m.

      End-tidal CO2 (ETCO2), referring to the level of the carbon dioxide released at the end of an exhaled breath, is required to be continuously monitored, especially in ventilated patients, as it is a sensitive and a non invasive technique that provides immediate information about ventilation, circulation, and metabolism functions. ETCO2 is normally lower than the arterial partial pressure and varies between 0.6 and 0.7 kPa.

      There are two methods used to measure carbon dioxide. The sidestream capnometer method samples gases at a set flow rate (150-200 mL/min) from a sampling area through small diameter tubing, and the mainstream analyser method that uses a direct measurement of the patient exhaled CO2 by a relatively large and heavy sensors. Sidestram method allows the analysis of multiple gases and anaesthetic vapours comparing to the mainstream method that does not allow the measurement of other gases.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 24 - A strain gauge operates on which of the following principles? ...

    Incorrect

    • A strain gauge operates on which of the following principles?

      Your Answer:

      Correct Answer: Varying resistance of a wire when stretched

      Explanation:

      The principle by which a strain gauge works is that when a wire is stretched, it becomes longer and thinner, and as a result, its resistance increases.

      A strain gauge, which is used in pressure transducers, acts as a resistor. When the pressure in a pressure transducer changes, the diaphragm moves, changing the tension in the resistance wire and thus changing the resistance.

      Changes in current flow through the resistor are amplified and displayed as a pressure change measure.

      A Wheatstone bridge, on the other hand, is frequently used to measure or monitor these changes in resistance.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 25 - A 240 volt alternating current (AC) socket from a wall is used to...

    Incorrect

    • A 240 volt alternating current (AC) socket from a wall is used to charge a direct current (DC) cardiac defibrillator. Name the electrical component that converts AC to DC.

      Your Answer:

      Correct Answer: Rectifier

      Explanation:

      There are two types of defibrillators
      AC defibrillator
      DC defibrillator

      AC defibrillator,
      consists of a step-up transformer with primary and secondary winding and two switches. Since secondary coil consists of more turns of wire than the primary coil, it induces larger voltage. A voltage value ranging between 250V to 750V is applied for AC external defibrillator. And used to enable the charging of a capacitor.

      DC defibrillator,
      consists of auto transformer T1 that acts as primary of the high voltage transformer T2. Is an iron core that transfers energy between 2 circuits by electromagnetic induction. Transformers are used to isolate circuits, change impedance and alter voltage output. transformers do not convert AC to DC.

      Diode rectifier composed of 4 diodes made of semiconductor material allows current to flow only in one direction. Alternating current (AC) passing through these diodes produces direct current (DC). Capacitor stores the charge in the form of an electrostatic field.

      Capacitor is used to convert the rectified AC voltage to produce DC voltage but capacitors do not directly convert AC to DC.

      Inductor induces a counter electromotive force(emf) that reduces the capacitor discharge value.

      In step-down transformer primary coils has more turns of wire than secondary coil, so induced voltage is smaller in the secondary coil.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 26 - Which one of the following statement is true regarding United Kingdom gas cylinders?...

    Incorrect

    • Which one of the following statement is true regarding United Kingdom gas cylinders?

      Your Answer:

      Correct Answer: Tensile tests are performed on sections of one cylinder in every hundred

      Explanation:

      Medical gas cylinders are made up of molybdenum steel but not cast iron. They are checked and assessed at a regular interval.

      At least one cylinder in each hundred are tested for tensile, pressure, smash, twist and straightening.

      Nitrous Oxide cylinders contain a mixture of liquid and vapour at a pressure of approx. 4500 kPa or 45 Bar. Carbon dioxide cylinder contain gas at the pressure of 5000kPa.

      The filling ratio is the ratio of mass of liquified gas in the cylinder to the mass of water required to fill the cylinder at the temperature of 15ÂșC. In the united kingdom, filling ratio of liquid nitrous oxide is 0.75. The cylinders are usually attached to the anaesthetic machine. As nitrous oxide is an N-methyl-d-aspartate receptor antagonist that may reduce the incidence of chronic post-surgical pain.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 27 - Bioelectric potentials that have been measured have an optimum bandwidth and typical frequency....

    Incorrect

    • Bioelectric potentials that have been measured have an optimum bandwidth and typical frequency. For a standard 12-lead ECG, which of the following bandwidth and voltage combinations is the best?

      Your Answer:

      Correct Answer: Bandwidth 0.05-150 Hz, voltage 100-4000 microvolts

      Explanation:

      The potential difference (amplitude) and bandwidth frequencies of bioelectric signals are typical.

      These are the following:

      ECG: A bandwidth of 0.5-50 Hz is usually sufficient in monitoring mode, but a typical diagnostic bandwidth is 0.05-150 Hz (up to 200 Hz) with a typical voltage range of 0.1-4 millivolts (100-4000 microvolts).
      EEG has a frequency range of 0.5-100 Hz and a voltage range of 0.5-100 microvolts.
      EMG has a frequency range of 0.5 to 350 Hz and a voltage range of 0.5 to 30 millivolts.

      Prior to display, these small signals will need to be amplified and processed further.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 28 - International colour coding is used on medical gas cylinders. Other characteristics also play...

    Incorrect

    • International colour coding is used on medical gas cylinders. Other characteristics also play a role in determining the gas's identity within a cylinder. Which of the following options best describes a cylinder containing analgesics for obstetrics?

      Your Answer:

      Correct Answer: Blue body, blue/white shoulder, full cylinder; 13700 KPa, gas mixture, requires a dual stage pressure regulator

      Explanation:

      The body of the Entonox cylinder is usually blue (occasionally white), with blue and white shoulders. Entonox contains a 50:50 mixture of oxygen and nitrous oxide, with a full cylinder pressure of 13700 KPa (137 bar). The cylinder is equipped with a two-stage pressure regulator for safe operation.

      The cylinder body and shoulder of nitrous oxide are (French) blue.

      In today’s anaesthetic workstations, carbon dioxide cylinders are no longer used.

      The body of an oxygen cylinder is black, with a white shoulder.

      The white Heliox (21 percent oxygen and 79 percent helium) cylinder has a brown and white shoulder. The administration of this gas mixture, which is less dense than air, is used to reduce turbulence (stridor) of inspiratory flow in patients with upper airway obstruction.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 29 - Regarding anaesthetic breathing circuits, which one would be the best for spontaneous breathing?...

    Incorrect

    • Regarding anaesthetic breathing circuits, which one would be the best for spontaneous breathing?

      Your Answer:

      Correct Answer: Lack circuit

      Explanation:

      Among the breathing circuits, the Lack circuit is the most efficient for spontaneous breathing.

      An outer coaxial tube is present to deliver fresh air; exhaust air is routed to an inner tube, which is then delivered to a scavenging system. An expiratory valve is seen at the patient end, which is an advantage over other circuits. Moreover, the Lack circuit prevents rebreathing slightly greater than the alveolar minute ventilation at 4-5 litres per minute.

      The Bain circuit prevents rebreathing at 160-200ml/kg per minute, and is a co-axial version of the Mapleson D circuit.

      The Mapleson E circuit prevent rebreathing at a fresh gas flow (FGF) of approximately twice the patient’s normal minute volume. A modification of this, the Mapleson F, has a reservoir bag at the opposite end for the FGF. This circuit is appropriate for paediatric patients with a body weight less than 20 kg.

    • This question is part of the following fields:

      • Anaesthesia Related Apparatus
      0
      Seconds
  • Question 30 - Prior to rapid sequence induction of anaesthesia, a man with a BMI of...

    Incorrect

    • Prior to rapid sequence induction of anaesthesia, a man with a BMI of 35 is pre-oxygenated. Which method of pre-oxygenation with a tight-fitting face mask is the most effective?

      Your Answer:

      Correct Answer: Oxygen 6 litres per minute via a Mapleson A breathing system, with patient sitting up at 30 degrees breathing four vital capacity breaths

      Explanation:

      This patient is morbidly obese and has a high risk of developing hypoxia. This will be exacerbated by the patient’s supine position, as a result of:

      Functional residual capacity has been reduced (FRC)
      Increased closing capacity (CC)
      Reduced tidal volume due to increased airway resistance, decreased thoracic cage compliance, and decreased respiratory muscle strength and endurance
      Following induction of general anaesthesia, there is a tendency for atelectasis and increased O2 consumption due to the increased workload of respiratory muscles and the overall increase in metabolism.

      Pre-oxygenation with 100 percent oxygen via a tight-fitting mask can be done using either tidal volume breaths for three to five minutes or four vital capacity breaths in normal circumstances. In the head-up position, this patient is much more likely to be adequately pre-oxygenated, maximising the FRC and minimising the CC. In spontaneously breathing patients, the Mapleson A and circle systems are both effective, but the Mapleson D requires 160-200 ml/kg/minute to prevent rebreathing.