A 53-year old female with a diagnosis of anaplastic thyroid carcinoma is admitted in the surgery department for an elective total thyroidectomy with radical neck dissection. The operation is expected to last for 10 hours. Which of the following is the most suitable humidifier to use in an anaesthetic circuit for this case?
Your Answer: Heat and moisture exchanger (HME)
Explanation:Adequate humidification is vital to maintain homeostasis of the airway. Heat and moisture exchangers conserve some of the exhaled water, heat and return them to inspired gases. Many heat and moisture exchangers also perform bacterial/viral filtration and prevent inhalation of small particles. Heat and moisture exchangers are also called condenser humidifier, artificial nose, etc. Most of them are disposable devices with exchanging medium enclosed in a plastic housing. For adult and paediatric age group different dead space types are available. Heat and moisture exchangers are helpful during anaesthesia and ventilatory breathing system. To reduce the damage of the upper respiratory tract through cooling and dehydration inspiratory air can be heated and humidified, thus preventing the serious complications. Moreover, they are the most appropriate humidification devices used for routine anaesthesia.
Gases can be bubbled through water to increase humidity. Passing gas through water at room temperature causes the gas to cool due to latent heat of vaporisation. The water bath can be heated. This improves the efficiency of the device and also reduces the incidence of bacterial colonisation.
Nebulisers use a venturi system which employs the Bernoulli effect. A gas at high flow passes through a constriction causing the gas to accelerate, reducing its potential energy allowing other gases or liquids to be entrained. This can include medications or in the case of humidification, water vapour. The size of the water droplet produced by nebulisation determines where in the airway it is deposited. Standard nebulisers produced droplets of 4 microns in diameter and these are deposited in the upper airway and trachea. Efficacy can be improved by passing the droplets over an anvil which further reduces particle size. The most efficient form of nebuliser is the ultrasonic nebuliser. Here a transducer immersed in water and vibrated at a frequency of 3MHz produces1-2micron droplets. These particles easily reach the bronchioles and provide excellent humidification.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 2
Incorrect
-
Regarding the carbon dioxide monitoring, which of the following statements is correct?
Your Answer:
Correct Answer: Carbon dioxide absorbs infrared radiation at 4.28 µm
Explanation:Carbon dioxide (CO2), is a carbonic gas made up of two dissimilar atoms, namely one carbon atom and two oxygen atoms. Capnography is a technique used to measure carbon dioxide during a respiratory cycle, and it consists in calculating the concentration of the partial pressure of CO2, through the absorption of the infrared light, namely that CO2 absorbs infrared radiation at a wavelength of 4.28 µm.
End-tidal CO2 (ETCO2), referring to the level of the carbon dioxide released at the end of an exhaled breath, is required to be continuously monitored, especially in ventilated patients, as it is a sensitive and a non invasive technique that provides immediate information about ventilation, circulation, and metabolism functions. ETCO2 is normally lower than the arterial partial pressure and varies between 0.6 and 0.7 kPa.
There are two methods used to measure carbon dioxide. The sidestream capnometer method samples gases at a set flow rate (150-200 mL/min) from a sampling area through small diameter tubing, and the mainstream analyser method that uses a direct measurement of the patient exhaled CO2 by a relatively large and heavy sensors. Sidestram method allows the analysis of multiple gases and anaesthetic vapours comparing to the mainstream method that does not allow the measurement of other gases.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 3
Incorrect
-
With respect to the peripheral nerve stimulators, which one is used to perform nerve blocking?
Your Answer:
Correct Answer: Using an insulated needle improves the success rate for the block
Explanation:The nerve stimulators deliver a stimulus lasting for 1-2 milliseconds (not second) to perform nerve blockage.
There are just 2 leads (not 3); one for the skin and other for the needle.
Prior to the administration of the local anaesthesia, a current of 0.25 – 0.5 mA (not 1-2mA) at the frequency of 1-2 Hz is preferred.
If the needle tip is close to the nerve, muscular contraction could be possible at the lowest possible current.
Insulated needles have improved the block success rate, as the current is only conducting through needle tip.
Stimulus to the femoral nerve which is placed in the mid lingual line causes withdrawer of the quadriceps and knee extension, that’s the dancing patella ( not plantar flexion).
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 4
Incorrect
-
From the following electromagnetic waves, which one has the shortest wavelength?
Your Answer:
Correct Answer: X rays
Explanation:Electromagnetic waves are categorized according to their frequency or equivalently according to their wavelength. Visible light makes up a small part of the full electromagnetic spectrum.
Electromagnetic waves with shorter wavelengths and higher frequencies include ultraviolet light, X-rays, and gamma rays. Electromagnetic waves with longer wavelengths and lower frequencies include infrared light, microwaves, and radio and televisions waves.
Different electromagnetic waves according to their wavelength from shorter to longer are X-rays, ultraviolet radiations, visible light, infrared radiation, radio waves. X-ray among electromagnetic waves has the shortest wavelength and higher frequency with wavelengths ranging from 10*-8 to 10* -12 and corresponding frequencies.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 5
Incorrect
-
Which of the following statements is TRUE regarding an epidural set?
Your Answer:
Correct Answer: 19G Tuohy needles have 0.5 cm markings
Explanation:A paediatric 19G Tuohy catheter is available that is 5cm in length and has 0.5cm markings
18G Tuohy catheters are generally 9 to 10cm to hub
Distal end of catheter is angled (15 to 30 degrees) and closed to avoid puncturing the dura
Epidural mesh are usually 0.2 microns and are used to filter bacteria and viruses to ensure sterility of procedure
Transparent catheters are 90cm long with diameters depending on gauge size. It has 1cm graduations from 5 to 20cm to ensure they have been inserted amply and removed completely. Distal end is smooth which can be open or closed (with lateral openings).
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 6
Incorrect
-
Which one of the following statement is true regarding United Kingdom gas cylinders?
Your Answer:
Correct Answer: Tensile tests are performed on sections of one cylinder in every hundred
Explanation:Medical gas cylinders are made up of molybdenum steel but not cast iron. They are checked and assessed at a regular interval.
At least one cylinder in each hundred are tested for tensile, pressure, smash, twist and straightening.
Nitrous Oxide cylinders contain a mixture of liquid and vapour at a pressure of approx. 4500 kPa or 45 Bar. Carbon dioxide cylinder contain gas at the pressure of 5000kPa.
The filling ratio is the ratio of mass of liquified gas in the cylinder to the mass of water required to fill the cylinder at the temperature of 15ºC. In the united kingdom, filling ratio of liquid nitrous oxide is 0.75. The cylinders are usually attached to the anaesthetic machine. As nitrous oxide is an N-methyl-d-aspartate receptor antagonist that may reduce the incidence of chronic post-surgical pain.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 7
Incorrect
-
From the following statements, which is true of humidification?
Your Answer:
Correct Answer: Ultrasonic humidifier can achieve greater than 100% relative humidity
Explanation:Increasing temperature increases the amount of water vapour contained in air; for example, at 20°C, air contains about 17 g/m3, and at 37°C, air contains about 44 g/m3. The wet and dry bulb hygrometer, like the hair hygrometer, measures relative humidity.
Under normal operating conditions, Heat and moisture exchangers (HMEs) allows relative humidity of up to 70% to be achieved. Mucus can impair their performance, and they should not be used for longer than 24 hours.
Hot water bath humidifiers might cause scalding, condensed water in the tubing can interfere with gas flow, and there is a danger of infection.
The ultrasonic humidifier operates at roughly 2 MHz and may attain relative humidity levels much above 100%.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 8
Incorrect
-
Which statement is correct about the Mapleson anaesthetic breathing circuits?
Your Answer:
Correct Answer: Mapleson A is most efficient for spontaneous ventilation
Explanation:Mapleson breathing system (or circuit) analysed five different arrangements of components of the breathing system:
Mapleson A – It is the most efficient for spontaneous respiration. The flow of fresh gas required is 70-85 ml/kg/min, i.e., approximately 5-6 lit./min fresh gas flow for an average adult.
Mapleson B and C – inefficient for both SV and PPV; requires gas flow of two to three times minute volume (100 ml/kg/min). Not commonly used but category C may be used for emergency resuscitation.
Mapleson D – efficient for PPV at gas flow equivalent to patient’s minute volume; the Bain’s circuit is a coaxial version of the Mapleson D
Mapleson E and F – for paediatric use; requires gas flow at two to three times the patient’s minute volume. The Mapleson F consists of an open-ended reservoir bag (Jackson-Rees modification). -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 9
Incorrect
-
All of the following statements about pH electrode are incorrect except:
Your Answer:
Correct Answer: A semi-permeable membrane reduces protein contamination
Explanation:Pulse oximeters combine the principles of oximetry and plethysmography to noninvasively measure oxygen saturation in arterial blood. A sensor containing two or three light emitting diodes and a photodiode is placed across a perfused body part, commonly a finger, to be transilluminated. Oximetry depends on oxyhaemoglobin and deoxyhaemoglobin, and their ability to absorb the beams of light produced by the light emitting diodes: red light at 660 nm and infrared light at 960 nm.
The isosbestic point is the point wherein two different substances absorb light to the same extent. For oxyhaemoglobin and deoxyhaemoglobin, the points are at 590 nm and 805 nm. These are considered reference points where light absorption is independent of the degree of saturation.
Non-constant absorption of light is often due to the presence of an arterial pulsation, whilst constant absorption of light is seen in non-pulsatile tissues.
Most pulse oximeters are inaccurate at low SpO2, but is accurate at +/- 2% within the range of 70% to 100% SpO2. All pulse oximeters demonstrate a delay in between changes in SaO2 and SpO2, and display average readings every 10 to 20 seconds, hence they are unable to detect acute desaturation episodes.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 10
Incorrect
-
Among the following, which statement is true regarding electrical safety in an operation theatre?
Your Answer:
Correct Answer: The higher the frequency of the current the less risk to the patient
Explanation:The operating theatre is an unusual place with several applications of electrical equipment to the human body. This can lead to potential dangers associated with it that need to be prevented. Electrical safety in the operation theatre is the understanding of how these potential dangers can occur and how they can be prevented.
Electricity can cause morbidity or mortality by one of the following ways:
(i) electrocution
(ii) burns
(iii) ignition of a flammable material, causing a fire or explosion.Electrocution is dependant on factors like duration of contact with electric current, the current pathway and the frequency and size of current.
Option A: The higher the frequency, the less effects of electrocution on the body.
Option B & D: Equipment can be classified in classes and types.
The class designation describes the method used for protection against electrocution. Class I is basic protection, class II is double insulation and class III is safety extra low voltage.
The type designation describes the degree of protection based on the maximum permissible leakage currents under normal and fault conditions.
Type B:
can be class I, II or III but the maximum leakage current must not exceed 100 µA. It is therefore not suitable for direct connection to the heart.
Type BF
Similar to type B, but uses an isolated (or floating) circuit.
Type CF
Only type CF protect against microshock as they allow leakage currents of 0.05 mA per electrode for class I and 0.01 mA for class II. Microshock is a small leakage current that can cause harm because of direct connection to the heart via transvenous lines or wires, bypassing the impedance of the skin, leading to ventricular fibrillation. Microshock current of 100 μA is sufficient to cause VF.Option C: A 75mA electrocution can cause ventricular fibrillation. Use the following as a general guide to understand the effect of current size on the body.
1 mA – tingling pain
5 mA – pain
15 mA – tonic muscular contraction
50 mA – respiratory muscle paralysis
75 mA – ventricular fibrillation.Option E: Wet skin reduces the resistance to current flow and therefore increases the effects of electrocution.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 11
Incorrect
-
A strain gauge operates on which of the following principles?
Your Answer:
Correct Answer: Varying resistance of a wire when stretched
Explanation:The principle by which a strain gauge works is that when a wire is stretched, it becomes longer and thinner, and as a result, its resistance increases.
A strain gauge, which is used in pressure transducers, acts as a resistor. When the pressure in a pressure transducer changes, the diaphragm moves, changing the tension in the resistance wire and thus changing the resistance.
Changes in current flow through the resistor are amplified and displayed as a pressure change measure.
A Wheatstone bridge, on the other hand, is frequently used to measure or monitor these changes in resistance.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 12
Incorrect
-
What feature is found in the flowmeters of modern anaesthetic machines ?
Your Answer:
Correct Answer: The use of stannic oxide increases the accuracy of flowmeters
Explanation:Flowmeters measure the rate at which a specific gas, that the flowmeter has been calibrated for, passes through. This calibration is done at room temperature and standard atmospheric pressure with an accuracy of +/- 2%.
Reading the flowmeter is done from the top of a bobbin (the midpoint of a ball). Oxygen is the last gas to be added downstream to the mixture delivered to the back bar as a safety feature. This prevents delivery of a hypoxic mixture.
Inaccurate flow measurements occur when the bobbin sticks to the inside wall of the flowmeter. Stannic oxide has been used as a successful antistatic substance thus, reducing the aforementioned risk.
Carbon dioxide being easily delivered is found on some older machines, but those attached flowmeters are limited by a maximum flow of 500 ml /min. Thus avoiding the delivery of a hypercarbic mixture.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 13
Incorrect
-
Regarding the classification of breathing systems, which of the following is true?
Your Answer:
Correct Answer: The Conway classification describes a functional classification based on whether a CO2 absorber is required
Explanation:Breathing system is an assembly of components which connects patient’s airway to anaesthesia machine through which controlled composition of gas mixture is dispensed. It delivers gas to the patient, removes expired gas and controls the temperature and humidity of the inspired mixture. It allows spontaneous, controlled, or assisted respiration. It may also provide ports for gas sampling, airway pressure, flow and volume monitoring.
Breathing systems have been classified by Conway and Mapleson.
Conway suggested a functional classification:
– Circuits requiring a CO2 absorber
– Circuits not requiring a CO2 absorberWilliam Mapleson designated varying arrangements of breathing system components (masks, breathing tubes, fresh gas flow inlets, adjustable pressure-limiting valves, and reservoir bags) as Mapleson A-E circuits.
Mapleson A: Arranged as FGF inlet, reservoir bag, APL valve, mask.
In this circuit, because the reservoir bag is between the FGF inlet valve and the APL valve, expired gas from the patient may re-enter the system and fill the reservoir bag during controlled ventilation. This is the most efficient system for spontaneous breathing as the FGF must only be equal to a patient’s minute ventilation to prevent rebreathing.Mapleson B: Arranged as reservoir bag, FGF inlet, APL valve, mask.
In this circuit, the FGF inlet is closer to the APL valve, which helps prevent the rebreathing concern in the Mapleson A circuit as above during controlled ventilation.Mapleson C: Arranged as reservoir bag, FGF inlet, APL valve, mask.
In this circuit, the arrangement is the same as the Mapleson B circuit. However, this circuit is shorter as it does not contain elongated corrugated tubing. This circuit also has the FGF inlet close to the APL valve to aid in preventing rebreathing.Mapleson D: Arranged as reservoir bag, APL valve, FGF inlet, and mask.
In this circuit, the arrangement interchanges the FGF inlet and APL valve of the Mapleson A circuit. This system prevents rebreathing by directing FGF towards the APL valve rather than towards the patient during exhalation.Mapleson E: Arranged as corrugated tubing, FGF inlet, and mask.
In this circuit, there is no reservoir bag and no APL valve. Given the inability to alter the pressure of the circuit, this is ideal for spontaneously ventilating neonates or paediatric patients where low-pressure ventilation is desired. The system prevents rebreathing, similar to the Mapleson D circuit.Jackson Rees later modified the Mapleson E by adding an open ended bag, which has since become known as the Mapleson F.
Mapleson F: Arranged as APL valve directly connected to reservoir bag, corrugated tubing, FGF inlet, and mask.
The system prevents rebreathing similarly to Mapleson D by directing FGF towards the APL valve. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 14
Incorrect
-
Do the medical gas cylinders in the UK have the following colour codes?
Your Answer:
Correct Answer: Oxygen cylinders have a black body with white shoulders
Explanation:The following are the colour codes for medical gas cylinder in the UK:
Oxygen cylinder has a dark body with white shoulders.
Nitrous oxide is French blue. Air encompasses a grey body with dark and white quarters on the shoulders.
Entonox contains a French blue body with white and blue quarters on the shoulders.
Carbon dioxide barrels are grey in colour.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 15
Incorrect
-
Which of the following statements is true about oxygen face masks?
Your Answer:
Correct Answer: The oxygen concentration delivered by high air flow oxygen enrichment devices is not dependent on the respiratory pattern of the patient
Explanation:The normal peak inspiratory flow in healthy individuals is 20-30 L/min during each normal tidal ventilation. This is expected to increase with greater respiratory rate and deeper inspiration.
Face masks are used to facilitate the delivery of oxygen from a breathing system to a patient. Face masks can be divided into two types: fixed performance or variable performance devices.
In fixed performance devices (also known as high air flow oxygen enrichment or HAFOE), fixed inspired oxygen concentration is delivered to the patent, independent and greater than that of the patient’s peak inspiratory flow rate (PIFR). No random entrainment is expected to occur at the time of PIFR, hence, the oxygen concentration in HAFOE devices is not dependent on the patient’s respiratory pattern.
Moreover, in HAFOE masks, the concentration of oxygen at a given oxygen flow rate is determined by the size of the constriction; a device with a greater entrainment aperture delivers a lower oxygen concentration. Therefore, a 40% Venturi device will have lesser entrainment aperture when compared to a 31% Venturi. Venturi masks allow relatively fixed concentrations of supplemental oxygen to be inspired e.g. 24%, 28%, 31%, 35%, 40% and 60% oxygen. These are colour coded and marked with the recommended oxygen flow rate.
Variable performance devices deliver variable inspired oxygen concentration to the patient, and is dependent on the PIFR. The PIFR can often exceed the flow rate at which oxygen or an oxygen/air mixture is supplied by the device, depending on a patient’s inspiratory effort. In addition, these masks allow expired air to be released through the holes in the sides of the mask. Thus, with increased respiratory rate, rebreathing of alveolar gas from inside the mask may occur.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 16
Incorrect
-
You are preparing to anaesthetize a 27-year-old woman for an acute diagnostic laparoscopy to rule out appendicitis. She has no medical history and does not take any medications on a regular basis. You're going to do a quick sequence induction. Which method of preoxygenation is the most effective and efficient?
Your Answer:
Correct Answer: Mapleson A circuit with a fresh gas flow of 100 ml/kg
Explanation:Professor Mapleson classified non-rebreathing circuits based on the position of the APL valve, which controls fresh gas flow.
The Mapleson A (Magill) circuit is most effective in spontaneous breathing, requiring only 70-100 ml/kg (the patient’s minute volume) of fresh gas flow. The patient inhales fresh gas from the reservoir bag and tubing during inspiration. During expiration, the patient adds dead space gas (gas that hasn’t been exchanged) to the tubing and reservoir bag in addition to the fresh gas flow. At the patient’s end, alveolar gas is vented through the APL valve. During the expiratory pause, the fresh gas flow causes more gas to be released.
The Mapleson A is inefficient during controlled ventilation. Venting occurs during inspiration rather than during the expiratory phase, as it does during spontaneous ventilation. As a result, unless a high fresh gas flow of >20 L/minute is used, alveolar gas is rebreathed.
During spontaneous ventilation, the Mapleson D circuit is inefficient.
The oxygen concentration in a Hudson mask is insufficient to allow for adequate pre-oxygenation.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 17
Incorrect
-
The equipment used for patient monitoring in theatre and intensive care settings have electrical safety requirements for the protection of hospital staff and patients. Of the different classes of electrical equipment listed, which is least likely to cause a patient to suffer a microshock?
Your Answer:
Correct Answer: II (CF)
Explanation:Microshock refers to ventricular fibrillation caused by miniscule amounts of currents or voltages (100-150 microamperes) passing through the myocardial tissue from external cables arising from electrical components within the cardiac muscle, for example, pacemaker electrodes or saline filled venous catheters.
The risk of shock changes with the construction of electrical equipment in question. The main classes of electrical equipment include: I: Appliances have a protective earth connected to an outer casing which prevents live elements from coming in contact with conductive elements. A fault in this equipment class will result in live elements coming in contact with the outer casing and allowing electrical flow into the protective earth. This triggers the protective fuse to disconnect the electric supply to the appliance.
II: These appliances have reinforced insulation. In the event of a fault which causes the first layer of insulation to fail, the second layer is able to prevent contact of live elements with outer casing.
III: These appliances have no insulation to provide safety, and rely solely on the use of separated extra low voltage source (SELV) which limits voltage to 25V AC or 60V DC allowing for a person to come in contact with it without risk of a shock under normal dry conditions. Under wet conditions, voltage supply should be lowered to reduce risk of shock. These devices have no risk of macroshocks, but some risk of microshocks.
Class I and II electrical appliances are further divided into subtypes developed to limit current leakage in the event of a singular fault:
B (body): Upper limit of current leakage is 500 µA. This current can cause skin tingling and microshocks, but is not sufficient to cause injury.
BF (body floating): These appliances have an isolating capacitor or transformer which separate the secondary circuit from the protective earth. The upper limit of current leakage is the same as type B.
CF (cardiac floating): Upper limit of leakage current during a singular fault is 50 microamps. It is least likely to result in a microshock. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 18
Incorrect
-
A 240 volt alternating current (AC) socket from a wall is used to charge a direct current (DC) cardiac defibrillator. Name the electrical component that converts AC to DC.
Your Answer:
Correct Answer: Rectifier
Explanation:There are two types of defibrillators
AC defibrillator
DC defibrillatorAC defibrillator,
consists of a step-up transformer with primary and secondary winding and two switches. Since secondary coil consists of more turns of wire than the primary coil, it induces larger voltage. A voltage value ranging between 250V to 750V is applied for AC external defibrillator. And used to enable the charging of a capacitor.DC defibrillator,
consists of auto transformer T1 that acts as primary of the high voltage transformer T2. Is an iron core that transfers energy between 2 circuits by electromagnetic induction. Transformers are used to isolate circuits, change impedance and alter voltage output. transformers do not convert AC to DC.Diode rectifier composed of 4 diodes made of semiconductor material allows current to flow only in one direction. Alternating current (AC) passing through these diodes produces direct current (DC). Capacitor stores the charge in the form of an electrostatic field.
Capacitor is used to convert the rectified AC voltage to produce DC voltage but capacitors do not directly convert AC to DC.
Inductor induces a counter electromotive force(emf) that reduces the capacitor discharge value.
In step-down transformer primary coils has more turns of wire than secondary coil, so induced voltage is smaller in the secondary coil.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 19
Incorrect
-
A 74-year-old with a VVI pacemaker is undergoing a hip replacement. Which of the following is most likely to predispose him to an electrical hazard?
Your Answer:
Correct Answer: Use of cutting unipolar diathermy
Explanation:A single chamber pacemaker was implanted in the patient. In VVI mode, a pacemaker paces and senses the ventricle while being inhibited by a perceived ventricular event. The most likely electrical hazard from diathermy is electromagnetic interference (EMI).
EMI has the potential to cause the following: Inhibition of pacing
Asynchronous pacing
Reset to backup mode
Myocardial burns, and
Trigger VF.Diathermy entails the implementation of high-frequency electrical currents to produce heat and either make incisions or induce coagulation. Monopolar cautery involves disposable cautery pencils and electrosurgical diathermy units. In typical monopolar cautery, an electrical plate is placed on the patient’s skin and acts as an electrode, while the current passes between the instrument and the plate. Monopolar diathermy can therefore interfere with implanted metal devices and pacemaker function.
Bipolar diathermy, where the current passes between the forceps tips and not through the patient and is less likely to generate EMI.
Whilst the presence of a CVP line may in theory predispose the patient to microshock, the use of prerequisite CF electrical equipment makes this very unlikely. The presence of a CVP line and pacemaker does not therefore unduly increase the risk of an electrical hazard.
Isolating transformers are used to protect secondary circuits and individuals from electrical shocks. There is no step-up or step-down voltage (i.e. there is a ratio of 1 to 1 between the primary and secondary windings).
A ground (or earth) wire is normally connected to the metal case of an operating table to protect patients from accidental electrocution. In the event that a fault allows a live wire to make contact with the metal table (broken cable, loose connection etc.) it becomes live. The earth will provide an immediate path for current to safely flow through and so the table remains safe to touch. Being a low resistance path, the earth lets a large current flow through it when the fault occurs ensuring that the fuse or RCD will quickly blow. Without an operating table earth, the patient is not at more risk of an electrical hazard because of the pacemaker.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 20
Incorrect
-
The Control of Substances Hazardous to Health (COSHH) regulations recommend air supply rates to specific environments. Which of the following statements is true?
Your Answer:
Correct Answer: Preparation rooms receive a volume of 0.1 m3 of air per second
Explanation:Control of Substances Hazardous to Health (COSHH) was established by government under the Health and Safety at Work act in 1989. Their employers work on identification and management of those substances that are dangerous to health. The implications for anaesthetists include gas scavenging, equipment contamination and environmental safety. Adequate ventilation is required in areas where anaesthetic gases are present. The minimum air supply that is legally required in each specific area is: Operating theatres: 0.65 m3/second. Anaesthetic rooms: 0.15 m3/s. Preparation rooms: 0.1 m3/s. Recovery rooms need 15 air changes per hour.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 21
Incorrect
-
Which of the following is correct for gas pipeline pressure?
Your Answer:
Correct Answer: 4 bar for oxygen
Explanation:Pipeline gases (in the UK this includes: Oxygen, Nitrous oxide, Medical air, and Entonox) are supplied at 4 bar (or 400 kPa), and compressed air is supplied at 7 bar for power tools.
Carbon dioxide and nitric oxide are usually only supplied in cylinders.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 22
Incorrect
-
A laser is a device that stimulates atoms or molecules to emit light at particular wavelengths and amplifies that light, typically producing a very narrow beam of radiation. This can be of visible, infrared, or ultraviolet wavelengths. They have been widely utilized in theatre environment. Which of the following safety measures is most likely to reduce chances of eye injury to the theatre personnel?
Your Answer:
Correct Answer: Wearing laser protective goggles
Explanation:Eye damage is the most common potential hazard associated with laser energy. Everyone in the laser treatment room has the risk of eye exposure when working with a Class 3b or Class 4 healthcare laser system, and damage to various structures in the eye depending on wavelength of the laser if they are unprotected.
Red and near-infrared light (400-1400 nm) has very high penetration power. The light causes painless burns on the retina after it is absorbed by melanin in the pigment epithelium just behind the photoreceptors.
Infrared radiation (IR), or infrared light (>1060 nm), is a type of radiant energy that’s invisible to human eyes and hence won’t elicit the protective blink.
Ultraviolet light (<400 nm) is also a form of electromagnetic radiation which is can penetrate the cornea and be absorbed by the iris or the pupil and cause burn injuries or cataract occur due to irreversible photochemical retinal damage.
Safety eyewear is the best method of providing eye protection and are designed to absorb light specific to the laser being used. Laser protective eyewear (LPE) includes glasses or goggles of proper optical density (OD). The lenses should not be glass or plastic. The LPE should withstand direct and diffuse scattered laser beams.
The laser protection supervisor (LPS) or LSO is an individual who is responsible for any clinical area in which lasers are used. They are expected to have a certain level of equipment and determine what control measures are appropriate, for each individual system, but their presence does not guarantee the chances of having an eye injury.
Class 1 lasers are generally safe under every conceivable condition and is not likely to cause any eye damage. Class 3b or Class 4 medical laser systems are utilized in healthcare which have their own safety precautions.
Polarized spectacles can make your eyes more comfortable by eliminated glare, however, they will not be able to offer any protection against wavelengths at which laser act.
Using short bursts to reduce energy is also not correct as it would still be harmful to eye. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 23
Incorrect
-
Dinamap is an automated blood pressure monitoring device. Which of these statements best fit its properties?
Your Answer:
Correct Answer: The cuff should be positioned at the same level as the heart
Explanation:Dinamap continuously measures the systolic, diastolic and mean arterial pressure along with pulse rate, thereby providing a continuous monitoring of the blood pressure using the osscillitonometric principle of measurement.
The device loses accuracy towards the extremes of BP and is more accurate with systolic compared with diastolic pressure. In arrhythmias such as AF, the devices are also inaccurate due to the major fluctuations associated with the individual pulse pressure variations.
The manual BP device is still the gold standard for BP measurement and monitoring.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 24
Incorrect
-
A current flows through a simple electric circuit. Which of the following electrical component configurations has the greatest potential difference?
Your Answer:
Correct Answer: Two 5 ohm resistors in series with a passing current of 10 ampere
Explanation:According to the Ohm’s law, the potential difference is defined as
V(potential difference) = I(current) x (R) resistance
So, for one resistor of 5 ohms, a 10 ampere current will generate:
V = I x R
V = 10 x 5
V = 50 voltsThe formula for resistors in series can be defined as:
R(total) = R1+R2
Hence, when a current of 10 amperes passes through two 5 ohms resistors that are connected in series, the potential difference is:
V = I x (R1+R2)
V = 10 x (5+5)
V = 10 x 10
V = 100 voltsThe formula for resistors that are connected in a parallel circuit is:
1/ Rtotal = 1/R1 + 1/R2
Hence, when a current of 10 amperes passes through two 5 Ω resistors that are connected in a parallel circuit, the potential difference is:
Rtotal = R1 × R2/ R1 + R2
Rtotal = 25/10
Rtotal = 2.5
V = I x R
V = 10 x (R1xR2 / R1 + R2)
V = 10 x (25/10)
V = 10 x 2.5
V = 25 voltsCapacitors are electronic components that have the ability to store energy and charge (Q). The derived SI unit of capacitance (C) is the farad (F), which is equivalent to one coulomb per volt (V). The typical capacitors usually have a very small capacitance range, that ranges from pico to microfarads. On the contrary, supercapacitors can have a capacitance of up to 1-5000 F.
There are a number of factors that eventually determine the capacitance (C). They are as follows:
– Larger plate area (A)
– Closer plate spacing (d)
– Permittivity (ε) of the material (dielectric) between the plates (vacuum<<<glass), – C = ε × A/dThe units of stored charge are coulombs (Q), which is equal to the pathway of one ampere of current per second.
Stored charge, capacitance and voltage can be defined by the following equation:V (potential difference across capacitor) = Q(charge) / C (capacitance)
In a parallel circuit, the formula of capacitors is:
Ctotal = C1 + C2
Hence, two 5 farad capacitors in a parallel circuit arrangement with a charge of 100 coulomb and capacitance 10 F will give a potential difference of::
V = 100/10
V = 10 voltsIn a series circuit, the formula for capacitors is:
1/Ctotal = 1/C1 + 1/C2
Hence, two 5 farad capacitors with a charge of 100 coulomb will give:
Ctotal = C1 × C2/C1 + C2
In the example total capacitance = 25/10 = 2.5 F
V = 100/2.5
V = 40 volts. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 25
Incorrect
-
Which is the most appropriate statement describing the function of flowmeters?
Your Answer:
Correct Answer: Constant pressure, variable orifice - Heidbrink flowmeters
Explanation:There are different models of flowmeters determined by the applied pressure and its orifice. For instance, the watersight flowmeter functions through applying variable pressure, and it has a variable orifice. In contrast, the bubble flowmeter is operated using a constant pressure and orifice. Flowmeters such as rotameters, Heidbrink and Peak have a constant pressure but variable orifice. On the other hand, flowmeters including a simple pressure gauge, water depression, and pneumotachograph have a constant orifice but variable pressure.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 26
Incorrect
-
Bacteria and viruses that are smaller than 0.1 μm in diameter can be filtered out using heat and moisture exchanger (HME) with a typical pore size 0.2 μm. Choose the most appropriate mechanisms of particle capture for most bacteria and viruses.
Your Answer:
Correct Answer: Diffusion
Explanation:Warming, humidifying, and filtering inspired anaesthetic gases is done by heat and moisture exchangers (HME) and breathing system filters. They are made of glass fibres materials and are supported by a sturdy frame. Pleating increases the surface area to reduce resistance to air flow and boost efficiency.
Filters’ effectiveness is determined by the amount and size of particles they keep out of the patient’s airway. The efficiency of filters might be classified as 95, 99.95, or 99.97 percent. Pores with a diameter of 0.2 µm are common. The following are examples of typical particle sizes:
Red blood cell – 5 µm
Lymphocyte – 5-8 µm
Viruses – 0.02-0.3 µm
Bacteria – 0.5-1 µm
Depending on particle size, gas flow speed, and charge, particles are collected via a number of processes. Mechanical sieve, interception, diffusion, electrostatic filtration, and inertial impaction are some of the options:Sieve:
The diameter of the particle the filter is supposed to collect is smaller than the apertures of the filter’s fibres.Interception:
When a particle following a gas streamline approaches a fibre within one radius of itself, it becomes attached and captured.
Diffusion:A particle’s random (Brownian) zig-zag path or motion causes it to collide with a fibre.
By attracting and capturing a particle from within the gas flow, it generates a lower-concentration patch within the gas flow into which another particle diffuses, only to be captured. At low gas velocities and with smaller particles (0.1µm diameter), this is more common.Electrostatic:
These filters use large diameter fibre media and rely on electrostatic charges to improve fine particle removal effectiveness.
Impaction due to inertia:
When a particle is too large to respond fast to abrupt changes in streamline direction near a filter fibre, this happens. Because of its inertia, the particle will continue on its original course and collide with the filter fibre. When high gas velocities and dense fibre packing of the filter media are present, this sort of filtration mechanism is most prevalent.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 27
Incorrect
-
Regarding laminar gas flow, which of the following options has the most influence on laminar flow?
Your Answer:
Correct Answer: Diameter of tube
Explanation:Laminar flow can be defined as the motion of a fluid where every particle in the fluid follows the same path of its previous particles. The following are properties of laminar flow of gas or fluids:
1. Smooth unobstructed flow of gas through a tube of relatively uniform diameter
2. Few directional changes
3. Slow, steady flow through straight smooth, rigid, large calibre, cylindrical tube
4. Outer layer flow slower than the centre due to friction, results in discrete cylindrical layers, or streamlines
5. Double flow by doubling pressure as long as the flow pattern remains laminarPoiseuille’s Law relates the factors that determine laminar flow. It indicates the degree of resistance to fluid flow through a tube. The resistance to fluid flow through a tube is directly related to the length, flow and viscosity; and inversely related to the radius of the tube to the fourth power. This means that, when the radius is doubled, there is increase in flow by a factor of 16.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 28
Incorrect
-
Regarding tracheal tubes, which of the following statements are true?
Your Answer:
Correct Answer: Uncuffed RAE tubes have two Murphy eyes
Explanation:Tracheal tubes are made of either disposable plastic or reusable red rubber.
The tube size refers to the internal diameter (ID) in mm which is marked on the outside of the tube (some manufacturers mark the external diameter on the outside).
Plastic tubes have a radiopaque line spanning the entire length of the tube, which allows their position to be identified on x-rays. The bevel located at the end of the tube is left-facing and oval in shape, which improves the view of the vocal cords during intubation.
Oxford tubes are L-shaped and have a bevel that faces posteriorly. They have thick walls that increase the external diameter, making for a wider internal diameter.
RAE (Ring, Adair, and Elwyn) tubes are preformed and can either be north or south facing and cuffed or uncuffed. The cuffed RAE tubes have one Murphy eye, whereas the uncuffed has two Murphy eyes. Uncuffed tubes are primarily used in paediatric anaesthesia and the two Murphy eyes ensure adequate ventilation- should the tube be too long.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 29
Incorrect
-
Prior to rapid sequence induction of anaesthesia, a man with a BMI of 35 is pre-oxygenated. Which method of pre-oxygenation with a tight-fitting face mask is the most effective?
Your Answer:
Correct Answer: Oxygen 6 litres per minute via a Mapleson A breathing system, with patient sitting up at 30 degrees breathing four vital capacity breaths
Explanation:This patient is morbidly obese and has a high risk of developing hypoxia. This will be exacerbated by the patient’s supine position, as a result of:
Functional residual capacity has been reduced (FRC)
Increased closing capacity (CC)
Reduced tidal volume due to increased airway resistance, decreased thoracic cage compliance, and decreased respiratory muscle strength and endurance
Following induction of general anaesthesia, there is a tendency for atelectasis and increased O2 consumption due to the increased workload of respiratory muscles and the overall increase in metabolism.Pre-oxygenation with 100 percent oxygen via a tight-fitting mask can be done using either tidal volume breaths for three to five minutes or four vital capacity breaths in normal circumstances. In the head-up position, this patient is much more likely to be adequately pre-oxygenated, maximising the FRC and minimising the CC. In spontaneously breathing patients, the Mapleson A and circle systems are both effective, but the Mapleson D requires 160-200 ml/kg/minute to prevent rebreathing.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 30
Incorrect
-
When the volume of gas is measured, its value should be supported by a suitable abbreviation. From the following options, what does the STPD stand for?
Your Answer:
Correct Answer: Standard Temperature and Pressure, Dry
Explanation:Gas is composed of large numbers of molecules moving in random directions, separated by distances. They undergo perfectly elastic collisions with each other and the walls of a container and transfer kinetic energy in form of heat. These assumptions bring the characteristics of gases within the range and reasonable approximation to a real gas, particularly how any change in temperature and pressure affect the behaviour of gas. According to different theories and laws proposed, mathematical equations are derived to calculate the volume of gas, also different abbreviations are being used according to given conditions. The abbreviations used are ATP, BTPS, and STPD.
ATP stands for ambient temperature and barometric pressure, it is used to describe the conditions under which volume of gas is measured.
BTPS stands for body temperature and pressure saturated with water vapor. These are conditions under which volume of gas exist and all results of lung volume determination should be quoted at BTPS.
STPD stands for standard temperature and pressure, dry (0C and 760 mm Hg). These are the conditions that are used to describe quantities of individual gases exchanged in the lungs. -