Question 1
Correct
-
Health workers are at increased risks of anaesthetic exposure. Therefore, The Control of Substances Hazardous to Health (COSHH) regulations require measures implemented to assess and control the risks related to this exposure. Among the following control measures, which one is most likely to limit potentially harmful exposure to anaesthetic agents within an operating theatre?
Your Answer: Total intravenous anaesthesia (TIVA)
Explanation:Exposure to anaesthetic hazards is one among the occupational exposures in manipulating toxic agents or inhaling toxic gases during anaesthetic practices.
Toxic gases mainly nitrous oxide, is one of the most gaseous anaesthetic agents that constitutes an important source of pollution. One of the safe and effective technics used in anaesthesia and reducing the amount of pollution is the Total Intravenous Anaesthesia (TIVA) which consists of using opioids in analgesia and propofol for the induction and the maintenance of anaesthesia. It refers to the administration intravenously of an anaesthetic, sedative, and/or tranquilizer. A less polluting but not the best way to get rid of the toxic aesthetic agents is the scavenger system that collects and expels the gas outside the medical environment. Yet, this technique still represents a hazard for the environment and still increase the risk of exposure for the health workers and clinical staff.
Fume cupboards are also not recommended to use because of their high pollution potency, mainly of the air resulting in a great harm for medical workers.
Supraglottic airways as well as the Air Changes per Hour technics could be harmful for both patients and health workers, increasing the risks of transmitted diseases, namely nosocomial infections.
Therefore, the Total Intravenous Anaesthesia technique (TIVA) is most likely to be safe and recommended to use.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 2
Incorrect
-
Regarding anaesthetic breathing circuits, which one would be the best for spontaneous breathing?
Your Answer: Mapleson E
Correct Answer: Lack circuit
Explanation:Among the breathing circuits, the Lack circuit is the most efficient for spontaneous breathing.
An outer coaxial tube is present to deliver fresh air; exhaust air is routed to an inner tube, which is then delivered to a scavenging system. An expiratory valve is seen at the patient end, which is an advantage over other circuits. Moreover, the Lack circuit prevents rebreathing slightly greater than the alveolar minute ventilation at 4-5 litres per minute.
The Bain circuit prevents rebreathing at 160-200ml/kg per minute, and is a co-axial version of the Mapleson D circuit.
The Mapleson E circuit prevent rebreathing at a fresh gas flow (FGF) of approximately twice the patient’s normal minute volume. A modification of this, the Mapleson F, has a reservoir bag at the opposite end for the FGF. This circuit is appropriate for paediatric patients with a body weight less than 20 kg.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 3
Correct
-
Regarding laminar gas flow, which of the following options has the most influence on laminar flow?
Your Answer: Diameter of tube
Explanation:Laminar flow can be defined as the motion of a fluid where every particle in the fluid follows the same path of its previous particles. The following are properties of laminar flow of gas or fluids:
1. Smooth unobstructed flow of gas through a tube of relatively uniform diameter
2. Few directional changes
3. Slow, steady flow through straight smooth, rigid, large calibre, cylindrical tube
4. Outer layer flow slower than the centre due to friction, results in discrete cylindrical layers, or streamlines
5. Double flow by doubling pressure as long as the flow pattern remains laminarPoiseuille’s Law relates the factors that determine laminar flow. It indicates the degree of resistance to fluid flow through a tube. The resistance to fluid flow through a tube is directly related to the length, flow and viscosity; and inversely related to the radius of the tube to the fourth power. This means that, when the radius is doubled, there is increase in flow by a factor of 16.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 4
Correct
-
When the volume of gas is measured, its value should be supported by a suitable abbreviation. From the following options, what does the STPD stand for?
Your Answer: Standard Temperature and Pressure, Dry
Explanation:Gas is composed of large numbers of molecules moving in random directions, separated by distances. They undergo perfectly elastic collisions with each other and the walls of a container and transfer kinetic energy in form of heat. These assumptions bring the characteristics of gases within the range and reasonable approximation to a real gas, particularly how any change in temperature and pressure affect the behaviour of gas. According to different theories and laws proposed, mathematical equations are derived to calculate the volume of gas, also different abbreviations are being used according to given conditions. The abbreviations used are ATP, BTPS, and STPD.
ATP stands for ambient temperature and barometric pressure, it is used to describe the conditions under which volume of gas is measured.
BTPS stands for body temperature and pressure saturated with water vapor. These are conditions under which volume of gas exist and all results of lung volume determination should be quoted at BTPS.
STPD stands for standard temperature and pressure, dry (0C and 760 mm Hg). These are the conditions that are used to describe quantities of individual gases exchanged in the lungs. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 5
Incorrect
-
Regarding a paramagnetic oxygen analyser, the following statements are TRUE:
Your Answer: Can measure gases dissolved in blood
Correct Answer: Utilises null deflection
Explanation:The electrons in the outer shell of an oxygen molecule are unpaired, thus it has paramagnetic properties and is attracted into a magnetic field.
It utilizes null deflection -True
Null deflection is a crucial principle in paramagnetic analysers (reflected beam of light on two photocells) which gives very accurate results (typically 0.1%).It can be used to measure the concentration of diamagnetic gases – False
Since most other gases are weakly diamagnetic they are repelled by a magnetic field (nitric oxide is also paramagnetic).Can measure gases dissolved in the blood – False
For accurate analysis the sample gas must be dried before passing into the analysis cell, for example, by passage through silica gel. Therefore, they are unsuitable to measure gases dissolved in blood.Does not require calibration – False
As with most measurement instruments paramagnetic analysers must be calibrated before use.E) The readings are unaffected by water vapour – False
Water vapour affects the readings hence for accurate analysis the sample gas must be dried before passing into the analysis cell, for example, by passage through silica gel. That is why they are unsuitable to measure dissolved blood gases. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 6
Incorrect
-
Which one of the following statement is true regarding United Kingdom gas cylinders?
Your Answer:
Correct Answer: Tensile tests are performed on sections of one cylinder in every hundred
Explanation:Medical gas cylinders are made up of molybdenum steel but not cast iron. They are checked and assessed at a regular interval.
At least one cylinder in each hundred are tested for tensile, pressure, smash, twist and straightening.
Nitrous Oxide cylinders contain a mixture of liquid and vapour at a pressure of approx. 4500 kPa or 45 Bar. Carbon dioxide cylinder contain gas at the pressure of 5000kPa.
The filling ratio is the ratio of mass of liquified gas in the cylinder to the mass of water required to fill the cylinder at the temperature of 15ÂșC. In the united kingdom, filling ratio of liquid nitrous oxide is 0.75. The cylinders are usually attached to the anaesthetic machine. As nitrous oxide is an N-methyl-d-aspartate receptor antagonist that may reduce the incidence of chronic post-surgical pain.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 7
Incorrect
-
Dinamap is an automated blood pressure monitoring device. Which of these statements best fit its properties?
Your Answer:
Correct Answer: The cuff should be positioned at the same level as the heart
Explanation:Dinamap continuously measures the systolic, diastolic and mean arterial pressure along with pulse rate, thereby providing a continuous monitoring of the blood pressure using the osscillitonometric principle of measurement.
The device loses accuracy towards the extremes of BP and is more accurate with systolic compared with diastolic pressure. In arrhythmias such as AF, the devices are also inaccurate due to the major fluctuations associated with the individual pulse pressure variations.
The manual BP device is still the gold standard for BP measurement and monitoring.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 8
Incorrect
-
When compared to unipolar diathermy, which of the following is more specific to bipolar diathermy?
Your Answer:
Correct Answer: Has a power output of up to 140 joules per second
Explanation:Electrocautery, also known as diathermy, is a technique for coagulation, tissue cutting, and fulguration that uses a high-frequency current to generate heat (cell destruction from dehydration).
The two electrodes in bipolar diathermy are the tips of forceps, and current passes between the tips rather than through the patient. Bipolar diathermy’s power output (40-140 W) is lower than unipolar diathermy’s typical output (400 W). There is no earthing in the bipolar circuit.
A cutting electrode and a indifferent electrode in the form of a metal plate are used in unipolar diathermy. The high-frequency current completes a circuit by passing through the patient from the active electrode to the metal plate. When used correctly, the current density at the indifferent electrode is low, and the patient is unlikely to be burned. Between the patient plate and the earth is placed an isolating capacitor. This has a low impedance to a high frequency current, such as diathermy current, and is used in modern diathermy machines. The capacitor has a high impedance to current at 50 Hz, which protects the patient from electrical shock.
High frequency currents (500 KHz – 1 MHz) are used in both unipolar and bipolar diathermy, which can cause tissue damage and interfere with pacemaker function (less so with bipolar diathermy).
The effect of diathermy is determined by the current density and waveform employed. The current is a pulsed square wave pattern in coagulation mode and a continuous square wave pattern in cutting mode.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 9
Incorrect
-
Regarding the Manley MP3 ventilator, which statement is true?
Your Answer:
Correct Answer: Is a minute volume divider
Explanation:It’s a minute volume divider – True
The Manley MP3 ventilator is classed as a minute volume divider. The entire fresh gas flow or minute volume is delivered to the patient, having been divided into readily set tidal volumes.Can efficiently ventilate patients with poor pulmonary compliance – False
Ventilating patients with poor pulmonary compliance is not easily achieved, which makes it an unsuitable ventilator for a modern ICU.Can generate tidal volume up to 1500ml – False
It can generate tidal volumes up to 1000 ml (not 1500 ml), and the inflation pressure can be adjusted by sliding a weight along a rail.Functions like a Mapleson A system during spontaneous ventilation – False
The ventilator functions like a Mapleson D breathing system (not Mapleson A) during spontaneous ventilation.Has three sets of bellows – False
The fresh gas flow drives the ventilator which allows rapid detection of gas supply failure. It has two sets of bellows (not three) and three unidirectional valves. -
This question is part of the following fields:
- Anaesthesia Related Apparatus
-
-
Question 10
Incorrect
-
Bioelectric potentials that have been measured have an optimum bandwidth and typical frequency. For a standard 12-lead ECG, which of the following bandwidth and voltage combinations is the best?
Your Answer:
Correct Answer: Bandwidth 0.05-150 Hz, voltage 100-4000 microvolts
Explanation:The potential difference (amplitude) and bandwidth frequencies of bioelectric signals are typical.
These are the following:
ECG: A bandwidth of 0.5-50 Hz is usually sufficient in monitoring mode, but a typical diagnostic bandwidth is 0.05-150 Hz (up to 200 Hz) with a typical voltage range of 0.1-4 millivolts (100-4000 microvolts).
EEG has a frequency range of 0.5-100 Hz and a voltage range of 0.5-100 microvolts.
EMG has a frequency range of 0.5 to 350 Hz and a voltage range of 0.5 to 30 millivolts.Prior to display, these small signals will need to be amplified and processed further.
-
This question is part of the following fields:
- Anaesthesia Related Apparatus
-